| Display | 4 digit LCD(Max.-999~2000), Character height:12.7mm | 
        
            | Indication Conversion Rate | Selected by 125ms, 250ms, 500ms, 1s, 2s, 5s and 10s (Initialization 1sec)*The faster the rate, the sooner the batteries will run out.
 | 
        
            | Accuracy | ±0.25%F.S.±1digit, ±0.35%F.S.±1digit(2MPa)±1%F.S.±1digit(100MPa) | 
        
            | Temperature Effect on Zero Balance | ±0.4%R.C./10°C(2~50MPa type) ±1%R.C./10°C(100MPa)
 | 
        
            | Temperature Effect on Load | ±0.4%R.C./10°C(2~50MPa type) ±1%R.C./10°C(100MPa)
 | 
        
            | Compensated Temperature Range (Temperature of medium to be measured)
 | -10°C~55°C(2~50MPa type) | 
        
            | Operaion Temperature Range | -20°C~70°C(2~50MPa type) -10°C~60°C(100MPa)
 | 
        
            | Zero Balance | ±1%R.C. | 
        
            | Safe Overload Rating | 150%(2~50MPa type) | 
        
            | 300%(Compound pressure type) | 
        
            | Durability | 10 million cycles or more | 
        
            | Battery Longevity | Approx. 3000H (Under our measurement condition)(*1) | 
        
            | Power | 4 single alkali dry cell (LR03): 2 Power ON/OFF via the front sheet keys
 Low battery warning: Battery icon lit up on the display
 | 
        
            | Functions | Tare(Auto zero), Peak hold, Maximum/Minimum monitoring facility, Function of moving average | 
        
            | Case Material | ABS | 
        
            | Pressure Port Material | SUS630(2~5MPa type) SUS630(100MPa) Equivalent to HastelloyC22 SUS316L(Compound pressure type)
 | 
        
            | Connection screw | R1/4 G1/2(100MPa) | 
        
            | IP Rating | Equivalent to IP65 | 
        
            | Dimensions | φ85×49.4 (D: Protrusions not included) | 
        
            | Weight | Approx. 200g (Batteries included) | 
        
            | Accessories | Operation manual (1 copy) (Specify English or Japanese.) Test report (1 copy), Test batteries (*2)
 |