■Suitable for measurement of viscous media since the flush type (no feed port) wetted part is hardly clogged up.
■Wetted part constructed integral with diaphragm does not use any sealed liquid, offering durability, reliability and long life.
■Abundant analog output available.
[4~20mA(2 wire system, 4 wire system), 1~5V, 0~5V, 0~10V]
Measured medium |
Water, oil, gases or other media that will not corrode SUS630. |
Pressure type |
Positive and compound pressures can be measured. |
Measured Range |
Can be specified within the range of 1 MPa to 50 MPa of rated capacity (Calibration: 0.5 MPa to 50 MPa).
Specify more than 50% of rated capacity as the calibration range. |
Output type |
Specify among 4 to 20 mA (2 wire system, 4 wire system), 1 to 5 V, 0 to 5 V, and 0 to 10 V. |
Specification
Non Linearity |
±0.5%R.C. |
Hysteresis |
±0.5%R.C. |
Compensated Temperature Range
(Temperature of medium to be measured) |
-10~50°C(No condensation or freezing) |
Safe Temperature Range
(Temperature of medium to be measured) |
-10~75°C(No condensation or freezing) |
Temperature Effect on Zero Balance |
±0.8%R.C./10°C |
Temperature Effect on Load |
±0.8%R.C./10°C |
Safe Overload Rating |
150%R.C. |
Ultimate Overload Rating |
200%R.C. |
Insulation Resistance |
100MΩ or more (DC50V) |
Pressure Port Material |
SUS630 or 15-5PH (Precipitation hardening stainless steels) (*1) |
Connection Screw |
G3/8 |
Fastening Torque |
10~20N・m |
Fastening Torque;Maximum |
30N・m |
Power |
Selectable: 4DC24V(±10%), 5DC12V(±10%)
*Only 24V DC available when analog output is A5. |
Cable |
4- or 2-core shielded (oilproof) cable 2m Standardφ6mm |
Weight |
Approx. 300g(Cable not included) |
Accessories |
Operation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy)
Aluminum packing(×1)(Copper packing × 1 for 50 MPa model only) |