Features


For direct temperature measurement on tanks and piping
Compact and small design
High measurement accuracy
Output signal: Pt 100, 3-wire technology or 4...20 mA, 2-wire technology
Process connections for food/ pharmaceutical/biotechnology

Yêu Cầu Giá Tốt

Thông tin sản phẩm
Constructional design / case
Mechanical design The Pt 100 is integrated directly into a thermowell. Various types of process connections are available.
Electrical connection circular connector with screw connection M12
Further electrical connections upon request.
Degree of protection per DIN EN 60529 IP
Temperature detecting element · thermowell Ø 6 mm; option: reduced tip Ø 4 mm, length see order code.
Upon request a calculation for thermowells can be made (for static or dynamic application) with certificate.
· flush mounted with PEEK insert for temperature decoupling (for G 1/2 B)
Measuring resistor 1 x Pt 100, tolerances per class A per DIN EN 60751
Output signal · Pt 100, 3- or 4-wire technology (internal connected, see connection diagram)
· 4…20 mA , 2-wire technology, programmable, option
· auxiliary power 8.5…36 V DC
· max. load (U–8.5 V)/0.023 A
· overrange 3.6…23 mA, infinitily adjustable
· error signal (following sensor breakage) 3.6…23 mA, infinitily adjustable
· damping time 0…30 s
· accuracy at 23 °C < 0.1 % of span
· temperature effect < 0.13 %/10 °C.
Further technical data see type series PA2430, data sheet T4-082-1.
Temperature ranges ambient temperature -40...+85 °C
process temperature -50...+200 °C
allowed storage temperature -40...+85 °C
Process connections designs see order code.
Gasket is not included !
Further process connections upon request.
Operating pressure max. 16 bar
exception: Varivent D=68 Code A1512
max. 10 bar
Materials wetted parts stainless steel mat.-no. 1.4404 (316L)
Hygienic design surface quality
surface roughness Ra ≤ 0.8 μm
electropolished upon request
Response time per DIN EN 60751, test procedure with flowing water (without transmitter)
· thermowell 6 mm: T 90 = 5.5 s
· thermowell with reduced tip Ø 4 mm: T 90 = 4.5 s