Reference cond.: |
per DIN EN 60770-1
TU = const. (15…25) °C
φ = const. (45...75) % r.F.
pU = const. (860...1060) mbar
UB = 24 V DC (± 3 V DC)
RB = 50 Ω, HART: 250 Ω
Ground connected
MBA = 0 bar |
Calibration position: |
Process connection bottom: vertical
Process connection back: horizontal |
Deviation of characteristic: |
Refer to the adjusted measuring span (Limit point method per DIN 16086) |
|
Nominal range 1-40 bar, 1-16 bar abs.:
Turndown 5:1 0.1 %
Turndown > 5:1 0.02 % x TD |
|
Nominal range 0.25 bar:
Turndown 5:1 0.15 %
Turndown > 5:1 0.03 % x TD |
Long-term drift: |
Refer to nominal range
≤ 0.1 %/year |
Operational availability: |
< 12 s |
Response time t90 at current output: |
for 20 Hz measuring rate: typically 120 ms
for 100 Hz measuring rate: typically 50 ms |
Temperature influence, case: |
Refer to nominal range |
|
Ambient temperature -20…80 C:
Nominal range 1-16 bar, 0.1 %/10K, max. 0.3 %
Nominal range 0.25 bar, 0.15 %/10K, max. 0.4 % |
|
Ambient temperature -40…-20 C:
Typical 0.2 %/10K |
Temperature-influence process-connection: |
Depending on design |
|
See data sheet for detailed information |