PPCH-G General Specifications
|
| Power Requirements |
85 to 264 VAC, 50/60 Hz, 75 W max |
| Temperature Range |
15 to 35 °C |
| Vibration |
Meets MIL-T-28800D |
| Weight (Typical) |
32 kg approx (70 lb) |
| Dimensions |
30 cm H x 52 cm W x 50 cm D (12 in. x 20.5 in. x 20 in.) with enclosure, also 6U H rack mount |
| Communications Ports |
RS232 (COM1, COM2), IEEE-488.2 |
| Operating Modes |
Gauge, absolute |
| Pressure Ranges |
Atmosphere to 100 MPa (15 000 psi) |
| Operating Media |
Nitrogen, air (others optional) |
| Drive Air Supply |
500 to 800 kPa (75 to 120 psi) |
| Test Gas Supply |
Greater than maximum pressure to be controlled by at least 5% of controller range, but not more than 15% greater than controller range, minimum 70% of controller range, ± 1 % stabillity. Flow adequate to maintain stable supply while slewing into test volume. |
| Pressure Connections |
Drive air supply: 1/8 in. NPT F Test gas supply: DH500 (equivalent to AE F250C, HIP HF4) Test: DH500 |
| Utility Sensor |
Precision/Resolution ± 0.10 % span / 0.001 % span Drivers (8) 12V, 1 A max total output CE Conformance Available, must be specified |