| Specifications |
| Product start-up time |
5 seconds minimum |
| Measurement period |
5–180 seconds |
| Reporting interval |
0–3600 seconds via RS232 communication |
| Principle of operation |
Laser Diode optical detection of actual particulates |
| International codes |
ISO 7 – 22, NAS 0 – 12 |
| Calibration |
By recognised online methods confirmed by the relevant ISO procedures: MTD – via a certified primary ISO 11171 automatic particle detector using ISO 11943 principles, with particle distribution reporting to ISO 4406:1996 |
| Recalibration |
Contact Parker Hannifin |
| Working pressure |
2–420 bar (30–6000 PSI) |
| Flow range through icountPDR |
Note : Flow may be bi-directional 40–140 ml/min (optimum flow 60 ml/min) (0.01 – 0.04 USGPM (optimum flow 0.016 USGPM)) |
| Online flow range via System 20 sensors |
Size 0 = 6 to 25 l/min (2–7 USGPM) Size 1 = 24 to 100 l/min (6–26 USGPM) Size 2 = 170 to 380 l/min (45–100 USGPM) |
| Ambient storage temperature |
–40°C to +80°C (–40°F to +176°F) |
| Environment operating temperature |
-30°C to +60°C (–22°F to 140°F) |
| Fluid operating temperature |
+5°C to +80°C (+41°F to 176°F) |
| Computer compatibility |
Parker recommends the use of a 9-way D-type connector. This can be connected to a USB port using a USB-serial adaptor. Note that these connectors/adaptors are NOT supplied with icountPDR units: contact Parker Hannifin for advice. |
| Moisture sensor calibration |
±5% RH (over compensated temperature range of +10ºC to +80ºC; +50ºF to +176ºF) |
| Operating humidity range |
5% RH to 100% RH |
| Moisture sensor stability |
±0.2% RH typical at 50% RH in one year |
| Power requirement |
Regulated 9–40Vdc |
| Current rating |
Typically 120mA |
| Certification |
IP69K rating |
| Analogue output options (specified when ordering) |
| Variable current |
4–20mA |
| Variable voltage |
0–5Vdc, 0–3Vdc (user selectable) |
| CAN-bus |
to SAE J1939 (e.g. Parker IQAN) |
| Moisture sensor |
Linear scale within the range 5% RH to 100% RH |