| Feature |
Specification |
| Principle of operation |
Laser diode optical detection of actual particulates |
| Dimensions |
H 530mm x W 210mm x D 410mm |
| Weight |
Approx 18kg |
| Operating temperature and humidity |
+5ºC to +60ºC (-41ºF to +140ºF)
20-85% RH (tested at 30ºC (86ºF), non-condensing) |
| Storage temperature and humidity |
-40ºC to +90ºC (-40ºF to +194ºF)
10-90% RH (tested at 30ºC (86ºF), non-condensing) |
| Moisture sensor calibration |
±5% RH (over a compensated temperature range of +10ºC to +80ºC (+50ºF to +176ºF) |
| Moisture sensor stability |
±2% RH typical at 50% RH in one year |
| International codes |
ISO 7 to 21, NAS 0 to 12, AS 0 to 12 |
| Contamination standards |
Refer to Parker ‘Guide to Contamination Standards’ (DD0000015) on CD
MTD : ISO 4406:1999; NAS 1638; AS4059E (Differential); AS4509E (Cumulative)
ACFTD : ISO 4406:1987; ISO4406:1991; NAS 1638; AS4509E (Differential);
AS4509E (Cumulative); GOST 17216 : 2001 |
| Channel sizes |
| Channel Sizes: MTD μm(c) |
| ISO 4406:1999 |
NAS 1638 |
AS4059E (cum) |
AS4059E (diff) |
MTD 8 Channel |
| >4 μm (c) |
4-6 μm (c) |
<4 μm (c) |
4-6 μm (c) |
>4 μm (c) |
| >6 μm (c) |
6-14 μm (c) |
<6 μm (c) |
6-14 μm (c) |
>6 μm (c) |
| >14 μm (c) |
14-21 μm (c) |
<14 μm (c) |
14-21 μm (c) |
>14 μm (c) |
| >21 μm (c) |
21-38 μm (c) |
<21 μm (c) |
21-38 μm (c) |
>21 μm (c) |
| >38 μm (c) |
38-70 μm (c) |
<38 μm (c) |
38-70 μm (c) |
>25 μm (c) |
| >70 μm (c) |
>70 μm (c) |
<70 μm (c) |
>70 μm (c) |
>30 μm (c) |
| - |
- |
- |
- |
>38 μm (c) |
| - |
- |
- |
- |
>70 μm (c) |
| Channel Sizes: ACFTD μm |
| ISO 4406:1999 |
NAS 1638 |
AS4059E (cum) |
AS4059E (diff) |
GOST 17216:2001 |
| >2 μm |
2-5 μm |
- |
- |
>2-5 μm |
| >5 μm |
5-15 μm |
<5 μm |
5-15 μm |
>5-10 μm |
| >15 μm |
15-25 μm |
<15 μm |
15-25 μm |
>10-25 μm |
| >25 μm |
25-50 μm |
<25 μm |
25-50 μm |
>25-50 μm |
| >50 μm |
50-100 μm |
<50 μm |
50-100 μm |
>50-100 μm |
| >100 μm |
>100 μm |
<100 μm |
>100 μm |
>100-200 μm |
|
| Calibration |
MTD : via a certified primary ISO 11171 automatic particle detector using ISO 11943 principles, with particle distribution reporting to ISO 4406:1996
ACFTD : fully traceable to gravimetric first principles |
| Recalibration |
Contact Parker Hannifin for advice |
| Fluid compatibility |
Mineral-based oils and petroleum-based fuel - Contact Parker Hannifin for advice |
| Fluid management |
Selectable on screen between 25ml to 100ml |
| Viscosity range |
1 to 300cSt |
| Working pressure |
3 bar maximum input pressure, if used on-line. Contact Parker Hannifin for further advice |
| Flow range through icountBSplus |
Test: 60ml/min |
| Connection interface (On Line) |
INLET: 6mm push-fit, DRAIN: 4mm push-fit |
| Fluid operating temperature (Oil) |
+5ºC to +70ºC (-41ºF to 176ºF) |
| Fluid operating temperature (Fuel) |
+20ºC to +70ºC (-4ºF to 158ºF) |
| Sample bottle size |
See Parker ACC Spares list. Contact Parker Hannifin for advice |
| Flush sample size |
Selectable option within the icountBS Software: 25ml to 100ml |
| Memory storage |
500 tests (Integrated Warning Level) |
| Printer |
Thermal dot line printer - see ACC spares list for replacement paper |
| Battery type |
Polymer Lithium Ion Battery pack (ACC6NW032) |
| Power requirements |
Intergrated supply into the icountBSplus unit |
| Certification |
CE Certified. Supplied with EC Declaration of Conformity Certificate |