| Max kV | 130 kV | 
        
            | Max. electron beam power | 10 W | 
        
            | X-ray source | Open transmission target tube | 
        
            | X-ray spot size | 3 µm | 
        
            | Defect recognition capability | 2 µm | 
        
            | Geometric magnification | 2.5x-2,400x | 
        
            | System magnification | Up to 36,000x | 
        
            | Imaging system (Standard) | 1.45 Mpixel 12-bit camera with 6" image intensifier | 
        
            | Imaging system (Option) | Varian 1313 or 2520 Digital flat panel1Mpixel 12bit camera with dual field 4"/6" image intensifier
 | 
        
            | Manipulator | 4-axis (X, Y, Z, T) | 
        
            | Rotate axis | Optional | 
        
            | Tilt | 0 - 75 degrees | 
        
            | Measuring volume: | Largest square in single map 406x406 mm (16x16")Absolute max 711x762 mm (28x30")
 | 
        
            | Max. sample weight: | 5 kg (11 lbs) | 
        
            | Dimensions (BxWxH) | 1,225 x 1,810 x 2,145 mm  (48x71x84")(incl. monitor bracket and lamp tower)
 | 
        
            | Weight | 1,935 kg (4,266 lbs) | 
        
            | Radiation safety | <1 μSv/hr at the cabinet surface | 
        
            | Control: | Inspect-X control and analysis software | 
        
            | Automation inspection | Optional | 
        
            | Computed Tomography | Optional | 
        
            | Applications | Real-time inspection of electronics(BGA, µBGA, flip-chip and loaded PCB boards)
 |