Specifications
|
Hot Block |
Cold Block |
Range |
50 °C to 670 °C (122 °F to 1238 °F) |
–30 °C to 140 °C (–22 °F to 284 °F) |
Accuracy |
± 0.2 °C at 50 °C
± 0.4 °C at 400 °C
± 0.65 °C at 600 °C |
± 0.25 °C (insert wells)
± 0.65 °C (fixed wells) |
Stability |
± 0.02 °C at 100 °C
± 0.06 °C at 600 °C |
± 0.02 °C at –30 °C
± 0.04 °C at 140 °C |
Uniformity |
± 0.2 °C (± 0.05 °C typical) |
± 0.05 °C (insert wells)
± 0.25 °C (fixed wells) |
Well Depth |
152 mm (6 in) |
124 mm (4.875 in) |
Heating Time to Max. |
30 minutes |
15 minutes |
Cooling Times |
120 minutes from 660 °C to 100 °C |
30 minutes from 140 °C to –30 °C |
Well Inserts |
1 interchangeable well accommodates multi-hole insert |
1 interchangeable well accommodates multi-hole insert, plus four outer wells, 1/4", 3/16", 3/16", and 1/8" |
Computer Interface |
RS-232 interface included with Model 9930 Interface-it control software |
Power |
115 V ac (± 10 %), 8.8 A or 230 V ac (± 10 %), 4.4 A, switchable, 50/60 Hz, 1150 W |
Size (HxWxD) |
292 x 394 x 267 mm (11.5 x 15.5 x 10.5 in) |
Weight |
16.4 kg (36 lb) |
NIST-Traceable Certificate
(8 points) |
Data at 50 °C, 100 °C, 200 °C, 300 °C,
400 °C, 500 °C, 600 °C, and 660 °C |
Data at –30 °C, 0 °C, 25 °C, 50 °C, 75 °C,
100 °C, 125 °C, and 140 °C |