Specifications
|
Parameter |
Value |
Temperature range |
−200 °C to 420 °C |
Nominal resistance at 0.01 °C |
100 Ω ± 0.5 Ω |
Temperature coefficient |
0.003925 Ω/Ω/°C nominal |
Calibrated Accuracy[1] (k=2)
|
± 0.012 °C at −200 °C
± 0.011 °C at 0 °C
± 0.028 °C at 420 °C |
Short-term repeatability[2] |
± 0.007 °C at 0.010 °C |
Drift[3] |
± 0.007 °C at 0.010 °C |
Hysteresis |
± 0.010 °C maximum |
Sensor length |
50.8 mm (2.0 in) |
Sensor location
|
9.5 mm ± 3.2 mm from tip (0.375 in ± 0.125 in) |
Sheath diameter tolerance |
± 0.08 mm (± 0.003 in) |
Sheath material |
INCONEL® 600 |
Minimum insulation resistance |
500 MΩ at 23 °C |
Transition junction temperature range[4] |
−50 °C to 150 °C (see footnote) |
Minimum immersion length[5]
(< 5 mK error) |
102 mm (4.0 in)
|
Maximum immersion length |
254 mm (10 in) |
Response time[5] |
8 seconds typical |
Self heating (in 0 °C bath) |
60 mΩ/°C |
Lead-wire cable type
|
PTFE-jacketed cable, PTFE insulated conductors, 24 AWG stranded, silver plated copper |
Lead-wire length |
182.9 cm ± 2.5 cm (72.0 in ± 1.0 in) |
Lead-wire temperature range |
−50 °C to 150 °C |
Calibration |
NIST-traceable calibration |
[1]Includes calibration uncertainty and 100 hr drift.
[2]Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)
[3]After 100 hrs at max temp, 95 % confidence (k=2)
[4]Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.
[5]Per ASTM E 644
|
Calibration Uncertainty
|
Temperature |
Expanded
Uncertainty (k=2) |
−197 °C |
0.012 °C |
−80 °C |
0.012 °C |
−38 °C |
0.011 °C |
0 °C |
0.009 °C |
156 °C |
0.011 °C |
230 °C |
0.013 °C |
420 °C |
0.021 °C |
Note: Laboratories may periodically reevaluate their uncertainties. Calibration uncertainties depend on the calibration process, the standards used, and the instrument performance. |
|