Specifications |
Temperature Range |
35 °C to 400 °C
(95 °F to 752 °F) |
Display Accuracy |
±0.5 °C to 200 °C
±1.0 °C to 400 °C |
Stability |
±0.2 °C to 300 °C
±0.3 °C to 400 °C |
Resolution |
0.01 ° |
Uniformity |
±0.3 °C at 100 °C
±0.6 °C at 200 °C
±0.9 °C at 300 °C
±1.4 °C at 400 °C |
Heating Time |
25 °C to 400 °C: 22 minutes |
Cooling Time |
400 °C to 100 °C: 65 minutes |
Stabilization Time |
8 minutes |
Controller |
Fluke Calibration Model 2200, microprocessor based, with RS-232 |
Readout |
°C or °F, switchable |
Sensor |
RTD, 100Ω |
Heater |
325-watt, solid-state controlled |
Surface Plate |
6061 aluminum; top surface machine finished to 0.0008 mm (0.000032 in), 96 mm (3.8 in) diameter accessible |
Power |
115 VAC (±10 %), 2.8 A or 230 VAC (±10 %), 1.4 A, specify, 50/60 Hz, 325 W |
Weight |
3.2 kg (7 lb.) with 2200 Controller |
NIST-Traceable Calibration |
Data at 50 °C, 120 °C, 190 °C, 260 °C, 330 °C, and 400 °C |