Features:
    - Ranges from 25 mbar to 700 bar
- Single or dual range configuration
- Pressure test and maintenance
- Transmitter calibration
- Loop set-up and diagnostics
- Switch verification
Applications:
    - Pressure test and maintenance
- Transmitter calibration
- Loop set-up and diagnostics
- Switch verification
The DPI 800 Series is a complete range of advanced, robust and simple to use hand-held instruments. Highly cost effective, these tools are ideal for test/calibration of many popular process parameters. Advanced features and technical innovations address more applications in less time and deliver results you can rely on.
 
Specifications: (Contact us for fully detail)
 
Pressure Test and Measurement
    
        
            | DPI 800 Pressure Indicator | The ideal tool for pressure test and measurement | 
        
            | PressureRanges | 10 in H2O to 10,000 psi (25 mbar to 700 bar) including vacuum options | 
        
            | All-Inclusive Accuracy | Can be relied on from one year to the next, even in tough environmental conditions (see specifications) | 
        
            | Dual Sensor Configuration | Extended measurement range and simultaneous two channel reading (P1 and P2 or P1 - P2) | 
        
            | Stainless Steel Sensor Construction | Available for compatibility with a wide range of fluids and gases (refer to range table) | 
        
            | Programmable Leak Test | Reports the pressure drop and leak rate | 
        
            | Advanced Features | Hold, maximum/minimum/average, alarm, and tare facilitate troubleshooting | 
    
 
Pressure Instrument and Loop Maintenance
    
        
            | DPI 802 Pressure Loop Calibrator | Provides simultaneous pressure and mA measurement for transmitter and loop maintenance | 
        
            | Dual Readings | Simultaneous measurement of pressure and mA for transmitter calibration and loop maintenance | 
        
            | 24V Loop Power Supply | Energizes transmitters and control loops | 
        
            | Automatic Switch Test | Captures open/closed trip values providing a fast and highly accurate “safety system” check | 
        
            | HART Resistor | Can be switched into the loop when required by a HART digital communicator and avoids the inconvenience of carrying a 250 O resistor | 
    
 
IDOS™ Flexibility
    
        
            | Intelligent Digital Output Sensor (IDOS) | Universal Pressure Modules are available from 10 in H2O to 10,000 psi (25 mbar to 700 bar). | 
        
            | Total Flexibility | IDOS modules can be used with any compatible instrument; for example, a DPI 812 RTD loop calibrator can become a fully featured pressure calibrator. | 
        
            | Plug and Play | Modules are interchangeable between instruments, requiring no set-up or instrument calibration. | 
        
            | Range Expansion | Achieved by adding modules (please refer to the IDOS Universal Pressure Module datasheet). | 
        
            | *S_Standard Accuracy | Total accuracy over 32°F to 122°F (0°C to 50°C), including one year stability and calibration uncertainty | 
        
            | *P_Premier Accuracy | Precision over 65°F to 82°F (18°C to 28°C) For operation over 41°F to 113°F (5°C to 45°C): 0.014% FS for ranges above 10 psi (700 mbar) 0.075% FS for ranges below 15 psi (1 bar)
 Stability over a year: 0.01% reading ranges above 5 psi (350 mbar) 0.03% reading ranges below 10 psi (700 mbar)
 Calibration uncertainty: 50 ppm of reading
 | 
        
            | Single or DualRange | One or two internal sensors can be selected. For dual range instruments, G/D ranges will be configured as G (gauge).Only one of the two sensors can be 1500 psi (100 bar) or above.
 | 
        
            | Overpressure (maximum transient/intermittent pressure) | 5 psi (350 mbar) and below 4 x FS 10 to 10,000 psi (700 mbar to 700 bar) 2 x FS
 Maximum working pressure: 1.1 x FS
 | 
        
            | Pressure Connections | 1/8 NPT female or G 1/8 female | 
    
 
DPI 800 Series Common Specification
    
        
            | Operating Temperature | 14°F to 122°F (-10°C to 50°C) | 
        
            | Storage Temperature | -4°F to 158°F (-20°C to 70°C) | 
        
            | Humidity | 0% to 90% non-condensing, Def Stan 66-31, 8.6 Cat III | 
        
            | Shock and Vibration | BS EN61010:2001, Def Stan 66-31, 8.4 Cat III | 
        
            | EMC | BS EN61326-1:1998 + A2:2001 | 
        
            | Safety | Electrical BS EN61010:2001. Pressure Equipment Directive (PED), Class SEP. CE marked | 
        
            | Display | Graphic LCD with backlight. Resolution 99999 | 
        
            | Size (l x w x h) and Weight | 7.1 in x 3.3 in x 2 in (180 mm x 85 mm x 50 mm), 18 oz (500 g) | 
        
            | Batteries | 3 AA alkaline, >50 hours measure, >10 hours 24V source |