| Technical Data LR-Cal LPC 200-T |
| Display |
large graphic display, with backlighting on/off switchable |
| Display resolution |
up to 6 digits, selectable |
| Temperature sensor |
Pt 100, 4-wire, -10...+50°C |
| Temperature accuracy |
±0.05°C |
| Pressure Measurement rate |
5 values per second |
| Functions |
Min-, Max-Memory, Tara, Pressure rate, Filter, Zeropoint-adjustment |
| Data logging |
up to 1000 values, cycle time adjustable 1...3600 seconds |
| Interface |
USB |
| Supply |
internal Lithium-Ion rechargeable battery; charging time <6 h. Alternative: Mains supply via charger |
| Battery life time |
approx. 25 h |
| Permissible medium temperature |
0...85% r.h. at 50°C, non-condensing |
| Permissible storage temperature |
-20...+60°C |
| Housing material |
Polyamide 12, Membrane keyboard, Transparent panels |
| Protection degree |
IP 65 (without connection ports) |
| Weight (without sensor) |
approx. 480 g |
| Dimension (without sensor) |
W 75/100 x H 50/82 x D 194 mm |
| CE marking |
Interference emission and immunity to EN 61326, declaration of conformity on request |
| Technical Data LPC 2 |
| Pressure port |
up to 1000 bar: 1/2" BSP male; >1000 bar: M16 x 11.5 female with sealing cone |
| Wettet parts |
Stainless steel |
| Permissible medium temperature |
-30...+105°C |
| Permissible storage temperature |
-40...+85°C |
| Housing material |
Stainless steel |
| Electricla connections |
Round plug, 8-pole, suitable for LR-Cal LPC 200, incl. connection cable |
| Weight |
approx. 220 g |
| Dimension |
diameter 40 x height 102.5 mm |
| CE marking |
89/336/EWG interference emission and immunity to EN 61326
97/23/EG pressure equipment directive (module H) |