             Range:  
             | 
            
             ±3000 volts, auto polarity. 
             | 
        
        
            | 
             Accuracy:  
              
             | 
            
             0.1%, ±0.003%/°C over +20 to +40°C 
             range (at recorder output). Useable to 
             +50°C. 
             | 
        
        
            | 
             Settling time:  
             | 
            
             <3½ms to 1% of 1kV step input. 
             | 
        
        
            | 
             Speed of 
            response: 
              
             | 
            
             for 2kV step: <3½ms (10% to 90%) 
            for 3kV step: <4ms (10% to 90%) 
            For typical input step: <3ms (10% to 90%) 
            for 1kV step: <2½ms (10% to 90%) 
             | 
        
        
            | 
             response:  
             | 
            
             typically ±3db to >300Hz. 
             | 
        
        
            | 
             Frequency                           
              
             | 
            
             Small signal frequency response 
             | 
        
        
            | 
             Noise:*  
              
             | 
            
             <0.3V rms or 2 volts peak-to-peak 
             wide band (bandwidth restricted 
             to<1kHz) referred to input.         
             | 
        
        
            | 
             Drift:*  
              
             | 
            
             <0.01V/hr after 1 hr warm-up (0.003V/ 
             hr typical). Not measurably affected by 
             10°C temperature variation or 
             changes between 10% and 90% 
             relative humidity. 
             | 
        
        
            | 
             Size:  
              
             | 
            
             4 x 8½ x 14 inches (10 x 21.8 x 35.6 
             cm), 1.75” rack mounting available (1 
             or 2 per rack). 
             | 
        
        
            | 
             Recorder  
            output:  
              
             | 
            
             Compressed analog output is input  divided by 1000 (optional Model 
             244AK has divide by 200) for external 
             loads greater than 1KΩ. Output 
             connector is BNC. 
             | 
        
        
            | 
             Surface resolution:         
              
             | 
            
             Determined by probe aperture size 
            resolution: and surface-probe separation. 
             Standard type 1017AE & AS probes 
             with 0.07” (1.75mm) aperture will 
             resolve a 0.10” (2.5mm) spot at 0.02” 
             (0.5mm) separation. 
             | 
        
        
            | 
             Weight:  
             | 
            
             7 lb. (3.2 kg) 
             | 
        
        
            | 
             Accessories Included 
             | 
            
             Manual on CD, line cord, phono plug 
              
             | 
        
        
            | 
             Power requirement:  
             | 
            
             100, 115, 230 VAC, ±10%, 50/60Hz, 15 watts. 
              
             | 
        
        
            | 
             Certifications:    
             | 
            
             CE Mark compliant 
             |