- Ideal for testing static-control
products
- Meets ESD Association ANSI/ESD
STM 11.11 and ANSI/ESD STM 11.12
plus ASTM D-257 and EIA-541
- Measures both resistivity and
resistance to ground
- Rugged and reliable for routine
factory use
- Fast and simple to use
- 8-hour rechargeable battery
- Selectable scientific notation or
3-place log readouts
- No adjustments required
Thông số kỹ thuật
Accuracy:
|
±0.1 decade thru 10E12, ±0.15 decade
above 10E12 (typical);
±0.1 decade thru 10E9, ±0.15 decade thru
10E11; ±0.2 decade above 10E12 (max)
|
8.0 x 103 Ω to 2.0 x 1012 Ω: 10 volts
|
1.1 x 105 Ω to 2.0 x 1013 Ω: 100 volts
|
Resistance range
|
: 8 x 103 to 2.0 x 1013 Ω
|
Display type:
|
16-char. alphanumeric LCD
|
Power:
|
Rechargeable battery with 100,
120, or 230 VAC adapter/
charger supplied
|
Information
shown:
|
Operating mode, measured
Ω or Ω/□ in log or scientific
notation, applied voltage,
low battery indication
|
Electrode Monroe Model 96101A-1
Type:
|
Guarded-ring-type designed to
ASTM D-257
|
Battery life:
|
Up to 8 hours continuous
operation per charge
|
Weight:
|
5 lbs. (2.3 kg)
|
Inner electrode
disc diameter:
|
1.2 inches (3.0 cm)
|
Xem thêm: