The OM-CP-IRTC101A features a 10 year battery life, 1 second reading rate, a multiple start/stop function, ultra-high speed download capability, 500,000 reading storage capacity, optional memory wrap, battery life indicator, optional password protection, programmable high-low alarms and more.
The OM-CP-IRTC101A was designed with our customers in mind. OMEGA offers free firmware upgrades for the life of the product so that data loggers already deployed in the field can grow with new technological developments. Units do not need to be returned to the factory for upgrades. The user can do this from any PC.
Thông số kỹ thuật
Multiple Start/Stop Mode Activation:
To Start the Device: Press and hold the pushbutton for 5 seconds,
the green LED will flash during this time. The device has started logging
To Stop the Device:
Press and hold the pushbutton for 5 seconds, the red LED will flash during this time.
The device has stopped logging.
Real Time Recording: May be used with PC to monitor and record data in real time
Alarm: Programmable high and low limits; alarm is activated when temperature reaches
or exceeds set limits
Cold Junction Compensation: Automatic, based on internal channel
Maximum Thermocouple Resistance: 3000Ω
Thermocouple Type: K
Range (°C): See To Order chart below
Sensor: ±2.0°C
Field of View: 60° (1:1)
Minimum Spot Size: 8 mm (0.3")
Spectral Response: 6.5 to 14 µ
Reading Rate: 1 reading every 2 seconds to 1 every 12 hours
Memory: 500,000 readings per channel (1,000,000 readings total)
software configurable memory wrap 250,000 readings in multiple start/stop mode
Wrap Around: Yes
DATA LOGGER INTERNAL TEMPERATURE CHANNEL
Range: -40 to 80°C (-40 to 176°F)
Temperature Resolution: 0.1°C (0.18°F)
Logger Calibrated Accuracy: ±0.25°C (±0.45°F)
REMOTE CHANNEL
Thermocouple Type: K (infrared, contact sales for other types)
Thermocouple Connection: Female subminiature (SMP)
Battery Type: 3.6V lithium battery included; user replaceable
Battery Life: 10 years typical at a 15 minute reading rate
Data Format: Date and time stamped °C, °F, K, °R, µV, mV, V
Time Accuracy: ±1 minute/month (at 20°C, stand alone data logging)
Computer Interface: USB (interface cable required); 115,200 baud
Software: XP SP3/Vista/7 and 8 (32 and 64-bit)
Operating Environment: -18 to 70°C (-40 to 158°F)
Dimensions: 36 x 56 x 16 mm (1.4 x 2.2 x 0.6")
Weight: 24 g (0.9 oz)
Infrared Thermocouple Dimensions: 3.25 L x 1.80 cm Dia. (1.28 x 0.71");
0.9 m (36") PFA coated unshielded stranded wire
OM-CP-WATERBOX101A
Operating Temperature: -20 to 80ºC (-4 to 176ºF)
IP Rating: IP65
Dimensions: 62 H x 93 W x 25 mm D (2.45 x 3.66 x 0.93")
Material: Anodized aluminum with polycarbonate cover
Weight: 140 g (5 oz)
Start Modes:
• Immediate start
• Delay start up to 18 months
• Multiple pushbutton start/stop
Stop Modes:
• Manual through software
• Timed (specific date and time)
Multiple Start/Stop Mode: Start and stop the device multiple times without having
to download data or communicate with a PC
LED Functionality:
Green LED Blinks: 10 second rate to indicate logging; 15 second rate to indicate delay start
Red LED Blinks: 10 second rate to indicate low battery and/or full memory; 1 second rate
to indicate an alarm condition
Password Protection: An optional password may be programmed into the device to restrict access
to configuration options. Data may be read out without the password.
Xem thêm: Thiết bị đo nhiệt độ hồng ngoại OS-500F, OS500C Omega