Standard pressure editions: Vacuum:-1 till 0 bar (Vacuum),Overpressure: 0.05 till 0.5 bar (Low),0.2 till 5 bar (Normal), 1 till 10 bar (High) and 2 till 15 bar (Extra high) |
Leak rate Limit: max:+/- 500 Pa min: 1 Pa/s |
Network connection: 100-240 VAC / 0.9 A 50-60 Hz |
No of display digits: 3 |
Weight: Approx. 16 kg |
Display sampling time: 100ms |
Size: 188mm x 400mm x deep 315mm |
Timer setting: 999.9 s |
Sealing grade:IP32 |
Hårdvara: 8 BIT CPU. |
Color: Silver gray (RAL 6021) with a dark green front panel. |
Test pressure source: Clean air 1 bar over test press. |
CE – marked |
Pilot pressure source: Clean air min 5 bar |
Designed for industrial use |
Ambient temperature: |
Differential pressure sensor: |
– Operating temperature: 5-40°C |
– Reading accuracy: 0.3 Pa |
– Storage temperature: -20-60°C |
– Sensor range: +/- 500 Pa |
Ambient humidity: |
– Sensor proof pressure: 1.5 MPa |
– 80% Relative Humidity |
Resolution: 1 Pa |
– 80% RH or less, no dew condensation |
Display units (all units available upon request): |
Communication: RS 232, TCP/IP (optional) |
– Test Pressure (standard): kPa, mBar |
Control I/O Port: DB-9P Male connector |
– Leak (stnd): Pa, mm3/s, 1/100 ml/min, ml/min, l/min |
Combination instruments available upon request |