Sentinel 500/501 Seaward

Mã sản phẩm: 500/501 |
Thương hiệu:

Giá bán: Liên hệ


Tên SP: Sentinel 500/501 Seaward

Mã SP: 500/501

Hãng: Seaward

Yêu Cầu Giá Tốt

Thông tin sản phẩm


AC Dielectric Strength

Programmable Voltage Range: 0.100 kV - 5.000 kV (2V/Step Resolution)
Frequency Independent of Supply: 50 or 60 Hz Selectable
Output Voltage Accuracy: ±1% + 5 Volts (No Load)
Voltage Display Resolution: 0.001kV
Current Measurement: 0.001mA - 100.0mA
Range and Accuracy: ±1.5% + 30 Counts when trip limit <1.11mA
±1.5% ± 3 Counts when trip limit >1.11mA
Current Display Resolution: 0.001mA / 0.001mA / 0.1mA
Selectable Range of Pass/Fail Levels: 0.001mA - 100.00mA
Maximum Rated Current: 100mA (0.5kV ≤ V ≤ 5kV)
10mA (0.1kV ≤ V ≤ 0.5kV)
Arc Detection: Yes
Ramp time: 0.1 - 999.9s
Test Time: Off / 0.5s - 999.9s

 

DC Dielectric Strength

Programmable Voltage Range: 0.100 kV - 6.000 kV
(2V/Step Resolution)
Frequency Independent 50 or 60 Hz Selectable of Supply: 50 or 60 Hz Selectable
Output Voltage Accuracy: ±1% + 5 Volts (No Load)
Voltage Display Resolution: 0.001kV
Current Measurement: 0.001mA - 20.0mA
Range and Accuracy: ±1.5% + 30 Counts when trip limit <1.11mA ±1.5% + 3 Counts when trip limit >1.11mA
Current Display Resolution: 0.001mA / 0.001mA / 0.1mA
Selectable Range of Pass/Fail Levels: 0.001mA - 20.00mA
Maximum Rated Current: 20mA (0.5kV ≤ V ≤ 5kV)
2mA (0.1kV ≤ V ≤ 0.5kV)
Arc Detection: Yes
Ramp time: 0.1 - 999.9s
Test Time: Off / 0.5s - 999.9s


Sentinel 501 Only

Earth Bond

Programmable Test Current: 03.00A - 32.00A ac
Test Current Resolution: 0.01A
Test Current Accuracy: 3A - 8A: (1% of rdg+0.2A)
8A - 32A: (1% of rdg+0.05A)
Test Voltage: 6V AC max (open circuit)
Test Voltage Frequency: 50Hz/60Hz selectable
Resistance Measurement Range: 10mΩ - 650.0mΩ
Resistance Measurement Resolution: 0.1mΩ
Resistance Measurement Accuracy: (1% of rdg + 2mΩ)
Test Time: 0.5 - 999.9s
Arc Detection: Yes

 

Insulation Resistance

Output Voltage: 50V - 1000V
Output-Voltage Resolution: 50V / Step
Output-Voltage Accuracy: ±(1% of setting +5V)[no load]
Resistance Measurement Range: 0.001GΩ -50.00GΩ
Test Voltage:  
50V - 450V 0.001 - 0.050GΩ
0.001 - 0.500GΩ
500V - 1000V 0.051 - 2.000GΩ
0.501 - 9.999GΩ
10.00 - 50.00GΩ
Selectable Range of Pass/Fail Levels: 0.001mA - 20.00mA
Output Impedance: Yes
RAMP (Ramp-Up Time): 600kΩ
TIMER (Test Time): 1s~999.9s


Sentinel 200

AC Dielectric Strength

Programmable Voltage Range: 0.100 kV - 5.000 kV (2V/Step Resolution)
Frequency Independent of Supply: 50 or 60 Hz Selectable
Output Voltage Accuracy: ±1% + 5 Volts (No Load)
Voltage Display Resolution: 0.001kV
Current Measurement: 0.001mA - 40.0mA
Range and Accuracy: ±1.5% + 30 Counts when trip limit <1.00mA
±1.5% ± 30 Counts when trip limit ≥1.00mA
Current Display Resolution: 0.001mA / 0.001mA / 0.1mA
Selectable Range of Pass/Fail Levels: 0.001mA - 40.00mA
Maximum Rated Current: 40mA (0.5kV ≤ V ≤ 5kV)
10mA (0.1kV ≤ V ≤ 0.5kV)
Arc Detection: Yes
Ramp time: 0.1 - 999.9s
Test Time: Off / 0.5s - 999.9s

 

DC Dielectric Strength

Programmable Voltage Range: 0.100 kV - 6.000 kV
(2V/Step Resolution)
Frequency Independent 50 or 60 Hz Selectable of Supply: 50 or 60 Hz Selectable
Output Voltage Accuracy: ±1% + 5 Volts (No Load)
Voltage Display Resolution: 0.001kV
Current Measurement: 0.001mA - 10.0mA
Range and Accuracy: ±1.5% + 30 Counts when trip limit <1.11mA ±1.5% + 3 Counts when trip limit >1.11mA
Current Display Resolution: 0.001mA / 0.001mA / 0.1mA
Selectable Range of Pass/Fail Levels: 0.001mA - 20.00mA
Maximum Rated Current: 10mA (0.5kV ≤ V ≤ 6kV)
2mA (0.1kV ≤ V ≤ 0.5kV)
Arc Detection: Yes
Ramp time: 0.1 - 999.9s
Test Time: Off / 0.5s - 999.9s

 

General Specification

Case Dimensions and Weight

Sentinel 500 330W x 140H x 482D mm
Approx 24kg
Sentinel 501 330W x 148H x 587D mm
Approx 27kg
Sentinel 200 330W x 150H x 460D mm
Approx 15kg

Part Number

Sentinel 500 489A920
Sentinel 501 489A921
Sentinel 200 489A910


SẢN PHẨM CÙNG DANH MỤC