| Frequency range |
300 kHz to 8 GHz ( R&S®ZVT8)
10 MHz to 20 GHz ( R&S®ZVT20) |
| Measurement time per test point |
|
| |
At 1 MHz IF |
<3.5 µs |
| Sampling time per test point |
|
| |
At 5 MHz IF (with R&S®ZVA-K17option) |
430 ns |
| Measurement time (201 test points) |
4.5 ms |
| Data transfer time (201 test points) |
|
| |
Via IEC/IEEE bus |
<2.9 ms |
| |
Via VX11 over 100 Mbit/s LAN |
<1.3 ms |
| |
Via RSIB over 100 Mbit/s LAN |
<0.7 ms |
| Switching time |
|
| |
Between channels |
<1 ms |
| |
Between instrument setups |
<10 ms |
| Dynamic range at 10 Hz measurement bandwidth, between test ports |
128 dB, typ. 138 dB ( R&S®ZVT8)
120 dB, typ. 133 dB ( R&S®ZVT20) |
| Output power at test port |
>13 dBm, typ. 15 dBm |
| Power sweep range |
>50 dB, typ. up to 60 dB |
| IF bandwidth |
|
| |
Standard |
1 Hz to 1 MHz |
| |
With R&S®ZVA-K17option |
5 MHz (10 MHz with restrictions) |
| |
With R&S®ZVA-K7option |
10 MHz (30 MHz with restrictions) |
| Number of channels, diagrams, traces |
>100 each |
| Number of test points per trace |
1 to 60001 |
| Operating system |
Windows XP Embedded |