Photochromic Lens Spectrophotometer BPC300 Bentham

Mã sản phẩm: BPC300 |
Thương hiệu: Bentham

Giá bán: Liên hệ

The BPC300 spectrophotometer offers accurate characterisation of photochromic (photosensitive) lens transmission in the faded and AM2 conditioned states.

Yêu Cầu Giá Tốt

Thông tin sản phẩm
Features


Monochromator based, collimated tuneable light source (280-780nm)
Temperature-controlled water bath with mount for plano lens sample
AM2 conditioning beam (including facility to attenuate and filter spectrum)
Optically chopped monochromatic beam with lock-in signal recovery enables measurement under AM2 activation
Automated reference level measurement
Fully-automated by Windows application including establishment and running of pre-defined measurement profiles
Generation of measurement report including parameters required by standards


Applications

ISO/ EN 8980-3:2013 - "Ophthalmic optics. Uncut finished spectacle lenses. Transmittance specifications and test methods."
ISO/ EN 12312-1:2013 - "Eye and face protection. Sunglasses and related eyewear. Sunglasses for general use."
ANSI Z80.3: 2015 - "Non-prescription Sunglass and Fashion Eyewear Requirements."
A S/NZS 1067-2016 - "Sunglasses and fashion spectacles."

Measurement


Spectral range of operation 280-780nm
Bandwidth 5nm
Wavelength accuracy ± 0.1nm (2400g/mm), ± 0.2nm (1200g/mm)
Beam diameter at sample plane 4mm diameter
Measured transmission range 0.5-100%
Maximum sample diameter 75mm
Conditioning beam AM2 with requirements of ISO8980-3:2013
0-50klx
Shutter
Optional filtering
Temperature range 5-50°C
Temperature accuracy ± 0.2°C
Bench space required 1m deep x 2.5m wide
Services requirements 6 x main sockets 1200W power consumption
Nitrogen gas supply where measurements below ambient are required


Monochromatic Probe Source

UV source 30W Deuterium lamp
UV-Vis source 100W Quartz Halogen lamp
Monochromator configuration Symmetric, single Czerny-Turner
Monochromator focal length 300mm
Diffraction gratings 2400 & 1200 g/mm
Diffraction grating mount Triple grating turret
Dispersion 5.4nm/mm (2400g/mm)
2.7nm/mm
Bandwidth 5nm
Wavelength accuracy ± 0.1nm (2400g/mm)
± 0.2nm (1200g/mm)
Chopping frequency 175 Hz (typical)
Beam diameter at sample plan 4mm diameter