Photobiological Safety Spectroradiometer IDR300-PSL Bentham

Mã sản phẩm: IDR300-PSL |
Thương hiệu: Bentham

Giá bán: Liên hệ
Test and certification companies and lamp and luminaire manufacturers can now evaluate the photobiological safety of lamps and lamp systems easier, faster and with greater accuracy than ever before.

Yêu Cầu Giá Tốt

Thông tin sản phẩm
Features


Double monochromator-based spectroradiometer (200-3000nm)
Motorised slits to vary instrument bandwidth throughout scan range
Irradiance : Precision transmission diffuser coupled to IDR300 with 1m long quartz fibre bundle
Radiance: Direct view telescope permitting measurement from 200mm to 10m in key 1.7, 5 and 11mrad FOVs. Coupled to IDR300 with 1m long quartz fibre bundle
Quartz halogen and deuterium irradiance standards, supplied with calibration traceable to PTB, Germany
Quartz halogen, integrating sphere-based radiance standard, supplied with calibration traceable to PTB, Germany
Beam profiler used to determine (apparent) source size and location
Photometric detector with virtual luxmeter operation for determination of illuminance (500 lx, dthr)
Benwin+ Windows spectroradiometer control
PSL Wizard measurement guidance, calculation and reporting application



Applications

Accurate photobiological safety evaluation to:

IEC 62471:2006, Photobiological safety of lamps and lamp systems
EN 62471:2008, Photobiological safety of lamps and lamp systems
JIS C 7550:2011, Photobiological safety of lamps and lamp systems
IEC 62471-5:2015, Photobiological safety of lamps and lamp systems - Part 5: Image projectors
EN 62471-5:2015, Photobiological safety of lamps and lamp systems - Part 5: Image projectors
IEC TR 62471-2:2009, Photobiological safety of lamps and lamp systems - Part 2
EU Low Voltage Directive, 2014/35/EU

Electro-Optical

Monochromator IDR300 double monochromator including integrated picoammeter and high voltage supply
Monochromator configuration Symmetric, Czerny-Turner
Monochromator focal length 600mm (200-1100nm, double mode)
300mm (1100-3000nm, single mode)
Bandwidth Software controlled motorised slit
Number of gratings 3 (Turret 1)
2 (Turret 2)
Resolution 0.075nm (2400g/mm)
0.3nm (1200g/mm)
0.9nm (400g/mm, single mode)
Dispersion 0.675nm (2400g/mm)
1.35nm/mm (1200g/mm)
8.1nm/mm (400g/mm)
Wavelength accuracy ± 0.05nm (2400g/mm)
± 0.1nm (1200g/mm)
± 0.6nm (400g/mm)
Detectors (200-1400nm) Multi-alkali photomultiplier
Silicon and InGaAs photodiodes
Detectors (1000-3000nm) PbS photodiode
Signal recovery (200-1400nm) 2 dual-channel picoammeters
Signal recovery (1000-3000nm) Pre-amplifier followed by lock-in amplifier
Irradiance entrance optic (200-1100nm) Cosine corrected transmission diffuser, 1m quartz fibre bundle
Irradiance entrance optic (1000-3000nm) Integrating sphere followed by CaF2 relay optic
Radiance entrance optic Direct view telescope permitting measurement from 200mm to 10m in key 1.7, 5 and 11mrad FOVs. Coupled to IDR300 with 1m long quartz fibre bundle
Spectral irradiance standards Quartz halogen and deuterium irradiance standards, supplied with calibration traceable to PTB, Germany
Spectral radiance standard Quartz halogen, integrating sphere-based radiance standard, supplied with calibration traceable to PTB, Germany
Beam profiler CMOS based beam profiler to determine (apparent) source size and location
Illuminance meter Photometric detector with virtual luxmeter