The removal of contaminants prior to cooling and distillation is essential to the cryogenic air separation process. If not detected quickly, impurities such as CO2 (carbon dioxide) can freeze in the downstream cryogenic equipment causing damage and product spoilage.

Yêu Cầu Giá Tốt

Thông tin sản phẩm
Applications


Air Separation Units & Nitrogen Plants
Gaseous CRMs & Calibration Gas Mixtures
High-Purity Gases & Systems
Hydrogen Fuel Cells
Research & Development
Syngas & Fuel Gas Production


Performance
Operating range See Detection Capability table
Detection limit (LDL, 24 h peak-to-peak) See Detection Capability table
Sensitivity (3σ) See Detection Capability table
Precision (1σ, greater of) ± 0.75% or 1/3 of Sensitivity
Accuracy (greater of) ± 4% or 1/2 of LDL
Speed of response < 3 minute to 95%
Environmental conditions 10°C to 40°C, 30% to 80% RH (non-condensing)
Storage temperature -10°C to 50°C
Gas Handling System and Conditions
Wetted materials 316L stainless steel (optional corrosion-resistant material), 10 Ra surface finish
Gas connections 1/4” male VCR inlet and outlet
Leak tested to 1 x 10-9 mbar l / sec
Inlet pressure 10 − 125 psig (1.7 − 9.6 bara)
Flow rate Up to 1.8 slpm
Sample gases Most inert, toxic, passive and corrosive matrices
Gas temperature Up to 60°C
Dimensions, H x W x D
Standard sensor 8.73" x 8.57" x 23.6" (222 mm x 218 mm x 599 mm)
Sensor rack (fits up to two sensors) 8.73" x 19.0" x 23.6" (222 mm x 483 mm x 599 mm)
Weight
Standard sensor 28 lbs (12.7 kg)
Electrical
Alarm indicators 2 user programmable, 1 system fault, Form C relays
Power requirements 90 − 240 VAC, 50/60 Hz
Power consumption 40 Watts max.
Signal output Isolated 4−20 mA per sensor
User interfaces 5.7” LCD touchscreen, 10/100 Base-T Ethernet, RS-232
  802.11g Wireless (optional), Modbus TCP (optional)