Moisture Sensor
(dew point sensor) |
- Range: -110°C to +20°C dewpoint ( -166°F to +68°F dewpoint )
- Accuracy: ±2°C temperature corrected
- Repeatability: 0.8°C
- Response time: 95% of step change in 3 min.
- Sample flow: >1 LPM
|
Temperature Sensor |
- Range: -40°C to +70°C
- Accuracy: ±2°C
|
Ergonomics |
- Easy and quick installation, no tools required
|
Electrical |
- 5VDC to 28VDC two wire loop operation 4mA to 20mA
- Housing isolated from 4-20mA Loop
- Measurement resolution of moisture sensor: 1pF resulting in less than 0.01 ̊C
- 5 pin M12 industrial connector IP68 when mating connector fully engaged
- Mating connectors with molded cables approved for Hazardous Location use available
- Two conductor cable min. 24AWG (for total cable length >1mile min. 20AWG) To meet CE requirements cable must be shielded.
- Built-in digital I/O for optionally controlling relays
|
Mechanical |
- All 316 Stainless Steel and PTFE wetted parts for fast response time.
- Enclosure all stainless steel construction IP67 protection when installed
- Dimensions (overall) including connector- L: 3.25″ (82.6mm) Dia: 1.62″ (41.3mm)
- Process threads 3/4″-16 and 14mm x 1.25mm Oring-Seal and 1/4″NPT, can install with either threads
- Weight Total: 6 oz (0.17 Kg)
- Pressure: 5000 PSIA (345 Bar)
|
Temperature Range |
- PLMa Analyzer: -20°C to +60°C
- Electronics: -40°C to +85°C
- Moisture Sensor: -20°C to +60°C
|
Miscellaneous Features |
- NIST traceable calibrations
- Units of measure: °C & °F dewpoint, ppmV, ppmW, µB H2O vapor pressure, grams of H2O / m3,
Lbs H2O /106 standard cubic feet in Natural Gas.
- Digital & Analog: Two Wire 4/20mA loop with HART®
- Transmitter stores all HART setup
|