DWX-15 |
DWX-15PD |
Applied voltage/applied steps:
*With 1 K Ω load |
1000 V - 15000 V
(200 V steps) |
Test impedance range |
50μH or higher |
Sampling speed/memory |
8bit/10nano-sec.(100MHz) / 8192-Byte |
Sampling ranges (WIDTH) |
-4、-3、-2、-1、0、1、2、3、4、5
10 ranges (0 is compatible with the DW Series) |
Test input circuit (Impedance) |
Resistance division (5MΩ) |
Screen display,
waveform display range |
640 x 460 (VGA), 8.4-inch TFT color liquid crystal, 4-color display
512 x 256 |
Judgment method |
Comparison of master waveform with test waveform Judgment based on comparison of waveform area, waveform differential area and discharge quantity Winding insulation dielectric breakdown voltage test function |
Master waveform memory capacity |
Body expansion |
196 types (14 types/14 pages)
700 types with Compact Flash Memory (14 types/50 pages) |
Microwave sensor antenna |
None |
Partial discharge electromagnetic sensing by narrow bandwidth flat batch antenna |
Detection range |
None |
It is recommended that a distance of 30 cm be allowed from the test piece to the antenna when the device is installed. |
Antenna dimensions/Length of cable between antenna and body
|
None
|
W = 80 mm, H = 80 MM, D = 10 mm, cable length = 2 m |
External interface |
Parallel I/O (Start, reset, OK, NG, Busy, mater switching, etc.)
RS-232C (Test control, test data, etc.)
Printer board (Hard screen copies, etc.)
Ethernet adapter (optional)
|
Accessories |
15-M test cable: 1, power supply cable (with 3P adapter)
Compact Flash Memory: 1, parallel I/O connector: 1
User’s manual, test result sheets: 1 set |
Usage conditions |
0℃ - 40℃, power supply voltage: 100 V 240 V ± 10% (Shipping configuration setting) |
External dimensions
(Not including carrying handle)
・Weight
|
W=364 x 517mm
H=517mm |