The model 2503A will accept inputs up to 10mA. Several ranges kV of 300, 150, 60, 30, 15, 6 & 3 are also built in to allow flexibility for various input voltages. On range 300 the full scale input is 360 kV, through a 50 pF low-loss high voltage standard capacitor while range 3 offers a full scale input of 3.6 kV. The full scale output is 120 volts.
Front panel and IEEE488 controllable. The two LCD displays monitor the null and output of the divider. The divider is housed in a rack mounted enclosure and is fully protected against transients. All connections are made to the rear of the instrument.
300 to 360,000 Volts
Outputs of 120 & 1V AC
Loss Measurement Systems
Power Calibration Systems
IEEE488 Interface
Division Accuracy < 20 PPM
Calibration Period 3 to 5 years