The FSL300 system provides accurate assessment of fibre spectral loss, paramount in quality control processes, material research and optical fibre development. Employing the internationally recognised reference test method, the cut-back technique, the FSL300 provides high measurement precision across all fibre types.

Yêu Cầu Giá Tốt

Thông tin sản phẩm

Features


Accurately assesses fibre spectral loss
Employs internationally recognised cut-back technique
The wavelength band may be extended throughout UV-vis and into the IR beyond 5000nm
Fully-automated via PC, responsible for dB/km loss calculation as a function of wavelength 
High-precision assessment of a large range of fibre types, including silica, doped glass, plastic and photonic bandgap fibres
Optical stability is ensured by the use of a quart tungsten halogen light source driven by a constant-current power supply
Light source and launch optics are bolted firmly to the monochromator to negate mechanical instability over time

FSL300_Si

Spectral range 350-1100nm
Typical spectral data interval 5nm
Typical spectral bandwidth (FWHM) 5nm
Light source Optically chopped IL1
Monochromator TMc300 with 1200 g/mm gratings
Detector 2.4x 2.4mm room temperature Silicon: 350-1100nm
Wavelength accuracy GA-T312R0U5 ± 0.2nm

FSL300_InGaAs

Spectral range 800-1700nm
Typical spectral data interval 5nm
Typical spectral bandwidth (FWHM) 5nm
Light source Optically chopped IL1
Monochromator TMc300 with 1200 g/mm gratings
Detector 1mm dia. room temperature InGaAs: 800-1700nm
Wavelength accuracy GA-T30831U2: ± 0.3nm

FSL300_InGaAs_EX

Spectral range 800-2500nm
Typical spectral data interval 5nm
Typical spectral bandwidth (FWHM) 5nm
Light source Optically chopped IL1
Monochromator TMc300 with 830 & 600 g/mm gratings
Detector 1mm dia. thermoelectrically cooled extended InGaAs: 800-2500nm
Wavelength accuracy GA-T30831U2: ± 0.3nm
GA-T306R1U6: ± 0.4nm

FSL300_PbSe

Spectral range 1500-5000nm
Typical spectral data interval 10nm
Typical spectral bandwidth (FWHM) 10nm
Light source ILD-QH-IR-CHOP
Monochromator TMc300 with 600 & 300 g/mm gratings
Detector 3mm dia. thermoelectrically cooled PbSe: 1000-5000nm
Wavelength accuracy GA-T306R1U6: ± 0.4nm 
GA-T303R3U0: ± 0.8nm