CT/PT Calibrator
Model: HEY-H
Hãng: Huatian

Yêu Cầu Giá Tốt

Thông tin sản phẩm

- Application of 320×240 LCD, broad vision, long life backlight, make it convenient to operate.
- All English interfaces, simple and nice.
- Extra—large display of percentage, ratio error and phase error, easy to observe.
- Automatic sampling, fully meet the requirements of JJG314—1993 and JJG314—1994.
- Automatic identification of polarity fault and ratio error mistake.
- Software can be upgraded with time. 
- Completely solve the testing problems that Class S encounters.
- Automatic switch of measuring range.
- Perfect combination of advanced circuit technique and DSP; ultimately avoid the instability of the RC phase –shift circuit.
- Power consumption:  <15VA (without micro-printer) <25VA (with micro-printer)
- Harmonic suppression ratio: >40db
- Dimensions: 260mm (L) ×350mm (W) ×150(H) mm
- Weight: 6kg

Thông số kỹ thuật

Working Condition

Temperature:5℃~40℃

Relative humidity:<80%(25℃时)

Altitude:<2500m

Power Frequency:50Hz±0.5Hz

Power Voltage:220V±5V

Measuring range

In-phase component (%):0.0001~200.0

Resolution:0.0001

Orthogonal Component (min): 0.001~700.0

Resolution:0.001

Impedance (Ω):0.0001~20.0

Resolution:0.0001

Admittance (ms):0.0001~20.0

Resolution:0.0001

Basic Error

In-phase Component

ΔX=±(X×2%+Y×2%)±Dx (with Class 1 to spare)

Orthogonal Component

ΔY=±(X×2%+Y×2%×34.48)±Dy (with Class 1 to spare)

“X”, “Y”——Values displayed

“Dx, Dy”——Quantization error

Dx=2,Dy=5

Percentage

Class 2 (with Class 1 to spare)

Working Range

Current

(1%~149%)In (In =5A)

(5%~149%)In (In =1A)

Voltage

(5%~149%)Un (Un =100V,150V,100V/)

(5%~149%)Un (Un =100V/3)

Working Load

Current

TO对TX<0.12Ω cosΦ=1

Voltage

a对x<0.25VA (100V)

Reversed Polarity Indication

Take action according to the polar instructions when the current is larger than 5% of the rated value and the error is larger than 180%.

Note: If no indication comes up when the current is larger than 10% of the rated value, malfunction exists. In that case please be sure not to increase the current so as not to burn the machine.

Indication of False Transformation Ratio

There should be an indication of false transformation ratio when the current is larger than 5% of the rated value and the error is between 30% and 180%.

Insulation and High Voltage Test

Terminal TX is connected with terminal ().

The power socket can bear a voltage of 1.5kV with respect to the outer shell for 1 min.