Cora 5X00 Raman Spectrometers Anton Paar

Mã sản phẩm: |
Thương hiệu: Anton Paar

Giá bán: Liên hệ

The Cora 5X00 series combines a small footprint – roughly the size of an open lab book – with the performance and high repeatability of a benchtop instrument. You can choose a single wavelength (Cora 5100, 5200 or 5300) or go for a dual-wavelength instrument (Cora 5500, 5600 and 5700) to ensure maximal Raman signal with minimal fluorescence when measuring different types of samples.

Yêu Cầu Giá Tốt

Thông tin sản phẩm


Single-wavelength Dual-band
Cora 5100 / Cora 5100 Fiber Cora 5200 / Cora 5200 Fiber Cora 5300 / Cora 5300 Fiber Cora 5500 / Cora 5500 Fiber Cora 5600 / Cora 5600 Fiber Cora 5700 / Cora 5700 Fiber
Optical
Excitation wavelength 532 nm 785 nm 1064 nm 532 nm and 785 nm 532 nm and 1064 nm 785 nm and 1064 nm
Spectral range 200 cm-1 to 3500 cm-1 100 cm-1 to 2300 cm-1 200 cm-1 to 3500 cm-1 for 532 nm 
100 cm-1 to 2300 cm-1 for 785 nm and 1064 nm
Resolution (FWHM) 9 cm-1 to 12 cm -1 6 cm-1 to 9 cm-1 12 cm-1 to 17 cm -1 9 cm-1 to 12 cm-1 for 532 nm 
6 cm-1 to 9 cm-1 for 785 nm 
12 cm-1 to 17 cm-1 for 1064 nm
Laser power 50 mW 0 mW to 450 mW (adjustable) 0 mW to 50 mW for 532 nm
0 mW to 450 mW for 785 nm and 1064 nm
Spectrograph f/2 Transmission Volume Phase Grating
Integration time 5 ms to 600 s 1 ms to 20 s 5 ms to 600 s for 532 nm and 785 nm 
1 ms to 20 s for 1064 nm
Wavelength calibration Automatic via software
Detector array 2048 px CCD 256 px InGaAs 2048 px CCD for 532 nm and 785 nm 
256 px InGaAs for 1064 nm
Detector cooling No cooling for 532 nm, approx. -5 °C for 785 nm and 1064 nm
Physical
Dimensions (DxWxH) 305 mm x 380 mm x 168 mm; 12 in x 15 in x 6.6 in
Weight ~ 6.4 kg (14 lb) ~ 7.3 kg (16 lb)
Operating ranges 0 °C to 40 °C; 0 % RH to 80 % RH
Electrical
A/D converter 16 bit
Power consumption <25 W <30 W
Battery (optional) Lithium ion
Power input 110 / 230 V AC