Typical measuring time |
3–5 minutes |
Maximum allowed volume of packages to be handled by PFD/PFD Plus |
Can: 0.5 L
Glass bottle: 1 L
PET bottle: 3 L |
Sample temperature |
15–25 °C (59–77 °F) |
Sample volume per measurement |
120–150 mL
(approx. 20 mL per additional measuring module) |
Compressed gas:
necessary quality of air if used as compressed gas supply
|
Class 5 from ISO 8573-1
max. particle size: 40 μm
max. pressure dew point: +7 °C (44.6 °F)
max. oil content: 25 mg/m³ |
Compressed gas:
necessary quality of nitrogen supply if Option O2 is installed: |
Purity ≥ 99.999 %
Class 5.0
max. particle size: 40 μm
max. pressure dew point: +7 °C (44.6 °F) |
Operating pressure for PBA system |
6 ±0.5 bar relative (87 ±7 psi) |
Dimensions (L x W x H) of the PBA system |
495 x 700 x 670 mm (19.5 x 27.6 x 26.4 in) |
Voltage |
AC 100 to 240 V, 50/60 Hz |
Power consumption |
190 VA (incl. external measuring modules) |
Power inlet |
according to IEC/EN 60320-1/C14, protection class I |
Fuses |
Glass tube fuses 5 x 20 mm; IEC60127-2; AC 250 V; T 5 AH |
Environmental conditions (EN 61010) |
Indoor use only |
Ambient temperature |
15–35 °C (59–95 °F) |
Air humidity |
10–90 % relative humidity, non-condensing |
Pollution degree |
2 |
Overvoltage category |
II |
Altitude |
max. 3000 m (9800 ft) |