| Principle | CO, CO2, CH4: Non-dispersion infrared (NDIR) sensor O2: galvanic sensor
 | 
        
            | Components | 1 component: CO, CO2, CH4 2 components: CO2 + CO, CH4 + CO, CO2 + CH4
 3 components: CH4 + CO2 + CO
 O2 measurement can be added to above all.
 | 
        
            | Range | CO: 0–200 ppm…100% CO2: 0–200 ppm…100%
 CH4: 0–1000 ppm…100%
 O2: 0–5% or 0–25%
 | 
        
            | Measurement range | 3 (Maximum range ratio 1 : 5) | 
        
            | Repeatability | ±0.5% FS | 
        
            | Zero drift | ≤ ±1.0% FS per day | 
        
            | Span drift | ≤ ±1.0% FS per day | 
        
            | Linearity | ±2.0% FS | 
        
            | Response time | ≤ 50 s (for 90% response) O2 measurement: ≤ 3 min
 | 
        
            | Output signal | 4–20 mA DC (linear) or 0–1 V DC (linear), up to 8 points, non-insulation (instantaneous value, O2-corrected instantaneous value, O2-corrected average value, CP calculated value)
 | 
        
            | Display | backlit LCD, English or Japanese | 
        
            | Functions | O2 correction, averaging, CP calculation, output signal hold, automatic display off | 
        
            | Communication | RS-232C (Modbus protocol) | 
        
            | Operating environment | 0–40°C, RH 90% or less | 
        
            | Power supply voltage | 100–115 V AC or 200–240 V AC, 50/60 Hz | 
        
            | Power consumption | Approx. 150 VA | 
        
            | Dimensions and weight | 365 (H) × 211 (W) × 527 (D) mm, approx. 12 kg | 
        
            | Gas conditions | Temperature | 0–40°C at inlet | 
        
            | Flow rate | 0.5 L/min ±0.2 L/min per optical system | 
        
            | Dust | ≤ 10mg/Nm3 | 
        
            | Pressure | ≤ 2 kPa (gas outlet should be open to the atmospheric air) | 
        
            | Mist | Unallowable | 
        
            | Corrosive gas | HCl: ≤ 10 ppm |