Simultaneous and continuous measurement of up to 4 components among: NO, SO2, CO, CO2, CH4, and O2

Ideal for flue gas monitoring in incinerators and industrial furnaces

Simultaneous and continuous measurement of up to 4 components
The mass flow type twin detector — hardly affected by interference gas
A variety of functions: auto calibration, O2-corrected output, averaging, alarm, etc.
Easy-to-see LCD
You can check the concentration changes of all components at a glance
19-inch rack mount
Built-in paramagnetic O2 sensor or optional zirconia O2 sensor

Yêu Cầu Giá Tốt

Thông tin sản phẩm

Specifications


Principle NO, SO2, CO, CO2, CH4: Non-dispersive infrared sensor
O2: built-in paramagnetic sensor or external zirconia sensor (ZFK7, optional) 
Components 1 component: NO, SO2, CO, CO2, CH4
2 components: NO + SO2, NO + CO, CO2 + CO, CH4 + CO, CO2 + CH4
3 components: NO + SO2 + CO, CO2 + CO + CH4
O2 measurement can be added to above all.
Range NO: 0–500 ppm…5000 ppm
SO2: 0–500 ppm…5000 ppm
CO: 0–200 ppm…100 vol%
CO2: 0–500 ppm…100 vol%
CH4: 0–1000 ppm…100 vol%
O2: 0–5 vol%…100 vol%
Repeatability ±0.5% FS
Zero drift ≤ ±2.0% FS per week
Span drift ≤ ±2.0% FS per week
Linearity ±1.0% FS
Response time ≤ 15 s (for 90% response in 2-component measurement)
Output signal • 4–20 mA DC (linear) or 0–1 V DC (linear)
(instantaneous values, O2-corrected instantaneous values, O2-corrected average values, O2 average) 
• Contact output 
(CO peak count alarm, concentration alarm, under auto calibration, device error, range identification, calibration error, etc.)
Contact input Auto-calibration start, average reset, range switch, output hold, etc.
Functions Auto calibration, O2 correction, averaging, CO peak count and alarm, output signal hold
Operating environment -5°C to +40°C, RH90% or less
Power supply voltage 100–240 V AC, 50/60 Hz, 70 VA
Dimensions 483 (W) × 177 (H) × 493 (D) mm
Gas conditions Temperature 0–50°C
Flow rate 1 L/min to 0.5 L/min
Dust ≤ 100 μg/Nm3 with particle size ≤ 0.3 μm
Pressure ≤ 10 kPa (gas outlet should be open to the atmospheric air)
Mist Unallowable
Corrosive gas ≤ 1 ppm
Moisture Below the saturation vapor density at room temperature (non condensing)