Thiết bị phân tích khí ZPA Fuji Electric

Mã sản phẩm: ZPA |
Thương hiệu: Fuji Electric

Giá bán: Liên hệ

Simultaneous and continuous measurement of up to 5 components among NO, SO2, CO, CO2, CH4, and O2 in incinerators or boilers


Compact and lightweight: 19-inch rack mount
Built-in magnetic O2 sensor or optional galvanic O2 sensor
Optional functions include: atmospheric pressure compensation, RS-485 communication, automatic calibration, alarm output, O2 correction, and more.
Low maintenance

Yêu Cầu Giá Tốt

Thông tin sản phẩm

Specifications

Principle Non dispersive infrared sensor (single beam system)
O2 sensor: built-in magnetic or galvanic sensor, or external zirconia sensor (ZFK7)
Ranges NO: 0–200 ppm…5000 ppm
SO2: 0–200 ppm…10 vol%
CO2: 0–100 ppm…100 vol%
CO: 0–200 ppm…100 vol%
CH4: 0–500 ppm…100 vol%
Built-in galvanic O2 sensor: 0–10 vol%…25 vol%
Built-in magnetic O2 sensor: 0–5 vol%…100 vol% or 100–95 vol%
External zirconia O2 sensor: 0–5 vol%…25 vol%
No. of ranges Up to 2 ranges per component
Repeatability ≤ ±0.5% FS
Linearity ≤ ±1% FS
Zero drift ≤ ±2% FS per week
For NO and SO2 of ranges below 500 ppm, auto calibration is necessary before measurement.
Span drift ≤ ±2% FS per week
Response time (for 90%) 10–30 s (depends on ranges)
Analog output signal 4–20 mA DC or 0–1 V DC
Display Backlit LCD (Japanese, English, or Chinese available) 
Instantaneous value, O2-corrected instantaneous value, O2 corrected average value, O2 average
Range switching by key operation, automatic, or by external contact input (option)
Contact input (option) Voltage contact input (12–24 V DC, up to 15 mA)
Remote range-switching, auto-calibration remote start, remote hold, average reset
Contact output (option) SPDT relay contact (24 V DC/1 A)
Device error, calibration error, range identification, auto-calibration status, solenoid valve drive for auto-calibration, limit alarm, CO peak alarm
Atmospheric pressure correction (optional) Built-in option available
Standard functions Output hold, auto/manual range switching
Optional functions Auto calibration, auto calibration remote start, remote hold of output, range identification contact output, high/low limit alarm, O2 correction, O2-corrected average values, average resetting contact input, CO peak alarm contact output
Communication (option) RS-485 (Modbus) (9-pin D-sub output): half-duplex, start-stop synchronization system
Gas inlet/outlet Rc 1/4 or NPT 1/4 internal thread
Purge gas flow rate 1 L/min (as needed)
Operating environment -5 to +45°C, RH90% or lower (non condensing)
Power supply voltage 100–240 V AC, 50/60 Hz
Power consumption Approx. 100 VA
Dimensions 133 (H) × 483 (W) × 382 (D) mm
Weight Approx. 11 kg
Conformity CE Marking

Sample gas conditions

Flow rate 0.5 L/min ±0.2 L/min
Temperature 0–50°C
Pressure ≤ 10 kPa
Dust ≤ 100 μg/Nm3 with particle size ≤ 0.3 μm
Mist Unallowable
Moisture Below the saturation vapor density at room temperature (non condensing)
Below the saturation vapor density at 2°C (non condensing) for NO, SO2, and CO of the range below 200 ppm
Corrosive gas HCl ≤ 1 ppm