Parameter |
Value |
Supply voltage 3-phase AC (50 Hz), V |
380 ± 38 |
Consumption Power, kV∙A, max |
50 |
Continuous operation at maximum capacity, minutes, max |
20 |
Pause after continuous operation, min, min
|
40
|
Power output, kVA |
400 |
Range of objects under the test capacities, µF |
0,01 - 1,0 |
Q-factor, min |
35 |
Present error of measurement of current and voltage, max, %
|
3 |
Operation threshold of protection current of the high voltage-side, A, max
|
13 |
Operation threshold of protection current of the low-voltage side, A, max
|
120
|
AC output voltage at test hydro, kV |
35 |
Dimensions L × W × H, mm
|
2500 × 2000 × 1800
|
Weight, kg, max |
3000 |