Control range max. |
5…95 %rh and 0…60 °C |
Control range min. |
5…95 %rh over the range of 10…50 °C |
Control stability |
<0.02°C, 0.1%rh |
Temperature gradient |
<0.05 (15...50 °C), <0.1 (5...60 °C), 0.15 (0 °C) |
Humidity gradient |
0.1 %rh or better |
Typical time to set-point |
5 minutes (35 to 80 %rh change, 23 °C) 10 minutes (23 to 45 °C change) |
Temperature sensor |
PT100 1/3 DIN Class B |
Humidity sensor |
Hygromer |
Control probe accuracy |
±0.8 %rh(10…95 %rh) / ±0.1 °C at 23°C |
Desiccant |
Molecular sieve |
Saturator |
Front panel fill, on-screen level indication |
Water quality |
Distilled or deionised water |
Water volume |
approx. 80 ml |
Chamber volume |
ca. 2 Liter |
External interface |
USB (9 ports), Ethernet (service use only) |
Power supply / Power consumption |
110...230 VAC, 50...60 Hz, 3 A |
External connections |
6mm temperature controlled sample loop connections |
Enclosure material |
Powder coated aluminium |
Dimensions |
450 x 406 x 205mm |
Weight |
13.0000 |