| Withstanding Voltage |
AC 3700V, 1 minute (Between input terminal and outer case) |
| Storage Temperature |
-10℃~60℃, 80% RH max (Without condensation) |
| Operating Temperature |
0℃ to 40℃, 80% RH max. (Without condensation) |
| Power supply |
1.5V (AM-3, R6 or "AA" size)×6 |
| Weight |
Approx. 330g (Excluding batteries) |
| Safety standard |
IEC 61010-1 (2001)
Installation categoryⅡ, 600V phase to earth. |
| Size |
170(W)×105(D)×54(H)mm |
| Accessories |
Line test lead:1
Earth test lead:1
Batteries:6
Test lead case:1
Belt:1
Instruction manual:1 |
| Optional accessory |
Test probe with remote switch |
| E.M.C. standard |
EN 61326 |
| Function |
Insulation resistance, AC voltage, Battery check |
| Meter movement |
100μA, 870Ω, Taut band meter |
| Constructional standard |
In accordance with IEC1557-2 or JIS C1302 (1994) |
| Overload protection |
120% of the highest nominal output voltage (10 sec.) |
| Battery check |
DC 6.3V~9.5V |
| Low battery limit |
DC 6.3V |
| Temperature characteristics(0~40℃) |
±5% rdg of specified accuracy |
| Accuracy(DC 50V) |
First effective range:0.01MΩ~5MΩ ±5%rdg、10MΩ~50MΩ ±30%rdg
Second effective range: |
| Accuracy(DC 125V) |
First effective range:0.02MΩ~10MΩ ±5%rdg
Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ ±10%rdg,10MΩ~100MΩ ±30%rdg |
| Accuracy(DC 250V) |
First effective range:0.05MΩ~20MΩ ±5%rdg
Second effective range:0.02MΩ~0.05MΩ 20MΩ~50MΩ ±10%rdg,50MΩ~100MΩ ±30%rdg |
| Accuracy(DC 500V) |
First effective range:0.1MΩ~50MΩ ±5%rdg
Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ ±10%rdg |
| Accuracy(DC 1000V) |
First effective range:2MΩ~1000MΩ ±5%rdg
Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ ±10%rdg |
| AC voltage (50/60Hz) |
Range/Accuracy/Input impedance/Maximum input voltage
AC 600V/±2.5% of full scale/Approx. 1.5M/AC 600V rms |