| Oakton pH/CON 700 General Specifications |
| Measures |
pH, mV, conductivity, total dissolved solids (TDS), and temperature (°C or F) |
| pHRange |
-1.99 to 16.00 |
| pHResolution |
0.01 |
| pHAccuracy |
±0.002 |
| pHCalibration |
Up to 5 points (USA: 1.68,4.01, 7.00, 10.01, 12.45 or NIST: 1.68,4.01, 6.86, 9.18, 12.45) with autobuffer recognition |
| mVRange |
±1999 mV |
| mVResolution |
0.1 mV (±199.9 mV) / 1 mV (beyond) |
| mVAccuracy |
±0.2 mV (±199.9 mV) / ±2 mV (beyond) |
| mVCalibration |
Offset up to ±150 mV |
| ConductivityRange |
0.00 ìS/cm to 200.0 mS/cm |
| ConductivityResolution |
0.01 ìS; 0.1 ìS ; 1 ìS ; 0.01 mS ; 0.1 mS |
| ConductivityAccuracy |
±1 % full scale |
| ConductivityCalibration |
Automatic (4 points); maximum 1 per range; manual (5 points); maximum 1 per range |
| TDSRange |
0.00 ppm to 100 ppt @ 0.5 TDS factor (to 200 ppt @ 1.0 TDS factor) |
| TDSResolution |
0.01 ppm; 0.1 ppm ; 1 ppm ; 0.01 ppt ; 0.1 ppt |
| TDSAccuracy |
± 1 % full scale |
| TDSCalibration |
Up to 5 points |
| TemperatureRange |
0.0 to 100.0 ºC / 32.0 to 212.0 ºF |
| TemperatureResolution |
0.1 ºC / 0.1 ºF |
| TemperatureAccuracy |
± 0.5 ºC / ± 0.9 ºF |
| TemperatureCalibration |
Off set in 0.1 º increments; off set range: ± 5 ºC / 9 ºF |
| Memory |
Up to 100 data sets |
| Dimensions |
6 1/8" x 6 7/8" x 2 3/4" (15.5 x 17.5 x 6.9 cm) |