Number of channels |
PicoScope 9341: 4
All other models: 2 |
Acquisition timing |
Selectable simultaneous or alternate acquisition |
Bandwidth, Full |
15 GHz |
20 GHz |
25 GHz |
Bandwidth, Narrow |
8 GHz |
10 GHz |
12 GHz |
Pulse response rise time, full bandwidth |
23.4 ps (10% to 90%, calculated) |
17.5 ps (10% to 90%, calculated) |
14.0 ps (10% to 90%, calculated) |
Pulse response rise time, narrow bandwidth |
43.8 ps (10% to 90%, calculated) |
35.0 ps (10% to 90%, calculated) |
29.2 ps (10% to 90%, calculated) |
Noise, full bandwidth |
< 1.2 mV RMS typical, < 1.6 mV RMS maximum |
< 1.5 mV RMS typical, < 2.0 mV RMS maximum |
< 1.9 mV RMS typical, < 2.5 mV RMS maximum |
Noise, narrow bandwidth |
< 0.7 mV RMS typical, < 0.9 mV RMS maximum |
< 0.8 mV RMS typical, < 1.1 mV RMS maximum |
< 1.0 mV RMS typical, < 1.3 mV RMS maximum |
Noise with averaging |
100 μV RMS system limit, typical |
Operating input voltage with digital feedback |
1 V p-p with ±1 V range (single-valued) |
Operating input voltage without digital feedback |
±400 mV relative to channel offset (multi-valued) |
Sensitivity |
1 mV/div to 500 mV/div in 1-2-5 sequence with 0.5% fine increments |
Resolution |
16 bits, 40 μV/LSB |
Accuracy |
±2% of full scale ±2 mV over nominal temperature range (assuming temperature-related calibrations are performed) |
Nominal input impedance |
(50 ±1) Ω |
Input connectors |
2.92 mm (K) female, compatible with SMA and PC3.5 |