            Oscilloscope specifications | 
            Specifications with PicoConnect 442 1000 V CAT III probe | 
        
        
            | Input channels | 
            4 channels | 
            4 channels | 
        
        
            | Analog bandwidth (–3 dB) | 
            20 MHz with D9 to BNC adaptors 
            15 MHz with PicoConnect 441 probe | 
            10 MHz | 
        
        
            | Rise time (calculated) | 
            17.5 ns with D9 to BNC adaptors 
            23 ns with PicoConnect 441 probe | 
            35 ns | 
        
        
            | Bandwidth limit | 
            100 kHz or 1 MHz (selectable) | 
            100 kHz or 1 MHz (selectable) | 
        
        
            | Vertical resolution, 12-bit mode | 
            12 bits on most input ranges 
            11 bits on ±10 mV range | 
            12 bits | 
        
        
            | Vertical resolution, 14-bit mode | 
            14 bits on most input ranges 
            13 bits on ±20 mV range 
            12 bits on ±10 mV range | 
            14 bits | 
        
        
            Enhanced vertical resolution 
            12-bit mode | 
            16 bits on most input ranges 
            15 bits on ±10 mV range | 
            16 bits | 
        
        
            Enhanced vertical resolution 
            14-bit mode | 
            18 bits on most input ranges 
            17 bits on ±20 mV range 
            16 bits on ±10 mV range | 
            18 bits | 
        
        
            | Input type | 
            Differential 
            9-pin D-subminiature, female | 
            Differential 
            9-pin D-subminiature, female | 
        
        
            | Input characteristics | 
            1 MΩ ±1%, in parallel with 17.5 pF ±1 pF (each differential input to ground). 
            < 1 pF difference between ranges. | 
            16.7 MΩ ±1%, in parallel with 9.3 pF ±1 pF (each differential input to ground) |