Magnetic measurement BH-5501 EMIC Japan

Mã sản phẩm: BH-5501 |
Thương hiệu: EMIC

Giá bán: Liên hệ

Yêu Cầu Giá Tốt

Thông tin sản phẩm

Feature:

Measures various kinds of materials from minute magnetic cores to permanent magnets.
The complete interactive software does not need trained skills.
The drift adjustment of the integrator is fully automatic.
The measuring speed is constant.
The latest digital feedback function added to the existing analogue feedback circuit makes it possible to measure test materials with the characteristics of radical rise of B accurately and easily.
All the measuring units are adopted SI unit.BH-5501 record DC magnetizating characteristics which is the basic characteristic of magnetic materials with high precision in a short period of time. And the data of various characteristics are computer-processed.
Our long years of experience developed the new model with high precision and easy operation.
Except for putting in and out the test materials, all operations and settings are made interactively through computer key-boards so that anyone can measure easily.


Measurement item

B-H Curve measurement (soft magnetic materials)

Maximum flux density (Bm), Residual flux density (Br), Maximum magnetic field (Hm),
Coercive force (Hc), Initial relative permeability (μi), Maximum permeability (μm), Squareness ratio (Br/Bm)

B-H (J-H) Curve measurement (Hard magnetic materials)

Maximum flux density (Bm), Residual flux density (Br), Maximum magnetic field (Hm), Coercive force (HCB),
Coercive force (HCJ), Maximum energy product ((BH)max), Squareness ratio (SR), HK, Magnetizing ratio (MG)

Measuring accuracy

B axis

±1%(Full-Scale ratio of the range)

H axis

±1%(Full-Scale ratio of the range)

Detection section

Magnetic field detection

Method : Hall element detection
Measuring sensitivity : ±40KA/m~±4000KA/m (4,8,16 7steps)

Excitation current detection

Method: non-inductive resistance Shunt detection
Measuring sensitivity : ±0.05A~±100A (1,2,5 11steps)

Flux detection

Method : CR integral detection
Measuring range : ±1×10-4  Wb to ±1×10Wb (6steps)

AD conversion

Sampling rate : 2m sec (max), Resolution : 16bit
Data memory : 130KB (per measurement)

Display

Display

15-inch LCD monitor

Contents

Measuring conditions, Characteristic values, B-H curve, J-H Curve

Printer

Ink-jet Paper size : A4, Printing time : 60 seconds

Power input

Detection section

AC100V 50/60Hz, about 1KVA

Power source section

Single-phase AC 200V 50/60Hz, about 2.4KVA

Power Source of Excitation
Current section※1

Method

Bipolar constant current method

Output

Output voltage ±50V (max) Output Current ±30A (max)