System Specifications |
Value |
Pump |
Flow rate range
Analytical gradient
Loading gradient (-TE model) |
5-50 µL/min or 20-200 µL/min (depends on model)
20-200 µL/min |
Maximum pressure 10,000 psi |
Maximum pressure 10,000 psi |
Retention time reproducibility |
< 0.5% RSD (@20 µL/min; 5-50 µL/min model) |
Gradient delay volume |
< 3 µL |
Autosampler |
Sample capacity |
6 micro titer plates (96 or 384 well)
6 vial adapters for 54 x 2 ml vial |
Injection volume range |
2-50 µL |
Injection volume precision |
< 1% RSD full loop
< 2% RSD partial loop |
Carryover |
< 0.005% (benzophenone) |
Injection/auxiliary valve |
6 port
6-32 thread size (1/32")
port-to-port volume < 60 nL |
Oven |
Temperature range |
+5°C to 60°C |
Inside dimensions (L x W x D) |
8" x 3/4" x 5/8" (20 x 1.9 x 1.6 cm) |
System |
Instrument control |
Controlled by SCIEX Analyst® 1.6.2 Software or higher, and TF 1.7 software or higher |
Dimensions (L x W x D) |
33.5" x 22.8" x 27.6" (85 cm x 58 cm x 70 cm) |