The SR715 LCR and SR720 LCR Meters measure passive components with as little as 0.05 % error. These easy-to-use instruments are quick to setup, adjust and calibrate. The SR715 and SR720 are ideal for applications such as incoming inspection, quality control, automated test, and general benchtop use.

Yêu Cầu Giá Tốt

Thông tin sản phẩm
Specifications  
Measurement modes Auto, R+Q, L+Q, C+D, C+R
Equivalent circuit Series or parallel
Parameters displayed Value, Deviation, % Deviation or Bin Number (Deviation and % Deviation are calculated relative to a stored value.)
Basic accuracy SR715: 0.2 %, SR720: 0.05 %
Measurement Range
(R+Q mode)
     R 0.0001 Ω to 2000 MΩ
     Q 0.00001 to 50
(L+Q mode)
     L 0.0001 µH to 99999 H
     Q 0.00001 to 50
(C+D mode)
     C 0.0001 pF to 99999 µF
     D 0.00001 to 10
(C+R mode)
     C 0.0001 pF to 99999 µF
     R 0.00001 Ω to 99999 kΩ
Electrical
Test frequency Fixed frequencies at 100 Hz, 120 Hz,1 kHz, 10 kHz (SR715 and SR720), 100 kHz (SR720 only)
Frequency accuracy ±100 ppm
Drive voltage Preset levels: 0.10, 0.25, 1.0 Vrms
Vernier: 0.1 to 1.0 Vrms with 50 mV resolution
Drive level accuracy ±2 %
Bias voltage Internal: 2.0 VDC ± 2 %
External: 0 to +40 VDC, fused @ 250 mA
Features
Averaging 2 to 10 measurements
Measurement rate Slow, Medium, Fast: 2, 10 or 20 measurements per second at test frequencies of 1 kHz and above and about 0.6, 2, 4 or 6 measurements per second at 100 Hz and 120 Hz.
Ranging Auto or Manual
Triggering Continuous, manual or remote over RS-232, GPIB or Handler interface
Binning Up to 8 pass bins, QDR and general fail bins, defined from the front panel or over the computer interfaces. Binning setup may be stored in non-volatile memory.
Fixture
Fixture 4-wire kelvin fixture for radial leaded parts with adapters for axial leaded parts.
Fixture protection Protected up to 1 Joule of stored energy and 200 VDC (for charged capacitors). Fused at 0.25 A output current for biased measurement.
Calibration
Zeroing Open and short circuit compensation
Compensation limits Short: R <20 Ω, Z <50 Ω, Open: Z >10 kΩ