FFT Analyzer SR760 and SR770 SRS Stanford Research System

Mã sản phẩm: SR760 and SR770 |
Thương hiệu: SRS Stanford Research System

Giá bán: Liên hệ
The SR760 and SR770 are single-channel 100 kHz FFT spectrum analyzers with a dynamic range of 90 dB and a real-time bandwidth of 100 kHz. Additionally, the SR770 includes a low-distortion source which allows you to measure the transfer functions of electronic and mechanical systems.

Yêu Cầu Giá Tốt

Thông tin sản phẩm
Frequency  
Measurement range 476 µHz to 100 kHz
Spans 191 mHz to 100 kHz in a binary sequence
Center frequency Anywhere within the 0 to 100 kHz measurement range
Accuracy 25 ppm from 20 °C to 40 °C
Resolution Span/400
Window functions Blackman-Harris, Hanning, Flat-Top and Uniform
Real-time bandwidth 100 kHz
Signal Input
Number of channels 1
Input Single-ended or differential
Input impedance 1 MΩ, 15 pF
Coupling AC or DC
CMRR (at 1 kHz) 90 dB (input range <-6 dBV)
80 dB (input range <14 dBV)
50 dB (input range ≥14 dBV)
Noise
     Typ. 5 nVrms/√Hz at 1 kHz (-166 dBVrms/√Hz)
     Max. 10 nVrms/√Hz (-160 dBVrms/√Hz)
Amplitude
Full-scale input range -60 dBV (1.0 mVp) to +34 dBV (50 Vp) in 2 dB steps
Dynamic range 90 dB (typ.)
Harmonic distortion No greater than -80 dB from DC to 100 kHz (input range 0 dBV)
Spurious No greater than -85 dB below full scale (<200 Hz). 
No greater than -90 dB below full scale to 100 kHz
(-50 dBV input range).
Input sampling 16-bit A/D at 256 kHz
Accuracy ±0.3 dB ± 0.02 % of full scale (excluding windowing effects)
Averaging RMS, Vector and Peak Hold. Linear and exponential averaging up to 64k scans.
Trigger Input
Modes Continuous, internal, external, TTL
Internal level Adjustable to ±100 % of input scale. Positive or negative slope.
Min. trigger amplitude 10 % of input range
External level ±5 V in 40 mV steps. Positive or negative slope. 10 kΩ Impedance
Min. trigger amplitude 100 mV
External TTL Requires TTL level, (low <0.7 V, high >2 V)
Post-trigger Measurement record is delayed by 1 to 65,000 samples (1/512 to 127 time records) after the trigger. Delay resolution is 1 sample (1/512 of a record).
Pre-trigger Measurement record starts up to 51.953 ms prior to the trigger. Delay resolution is 3.9062 ms.
Phase indeterminacy <2°
Display Functions
Display Real, imaginary, magnitude or phase
Measurements Spectrum, power spectral density, time record and 1/3 octave
Analysis Band, sideband, total harmonic distortion and trace math
Graphic expand Display expand up to ×50 about any point
Harmonic marker Displays up to 400 harmonics
Data tables Lists Y values of up to 200 points
Limit tables Detects data exceeding up to 100 user-defined upper and lower limit trace segments.
SẢN PHẨM CÙNG DANH MỤC