Input Sensitivity |
0.05V |
Dose/CPM Switchable |
External Dose / CPM Switch |
Display-LCD |
CPM, KCPM, CPS, uR/hr,mR/hr,R/hr,uSv/hr,mSv/hr,Sv/hr |
LCD Refresh Rate |
0.8 - 3.0 Sec adjustable |
Display Backlight |
push button, 30 Second auto off |
Scaler Count Time Ranges |
10-2550 Seconds (10 second increments) |
Probe 1 Ranges |
0-999.9 R/hr 0-999.9 Kcpm |
Probe 2 Ranges |
0-999.9 R/hr 0-999.9 Kcpm |
Scaler Ranges |
0-9,999 Kcpm |
Alarms (optional) |
5-95% of full scale for Dose & CPM |
Diagnostics |
Low Battery, Overrange - Audible and Visual Alarms |
Over Range Alarm |
Adjustable 10% above max range - dose calibration only |
Electrical Non-Linearity |
< 5% |
Response Time |
Fast - 10-2550 Seconds (adjustable) | Slow 10-250 seconds (adjustable) |
Drift |
<5% |
Environmental |
-20° F[-28° C]-140° F[60° C] | 5-95% RH IPX4 Housing |
#1 High Voltage |
Adjustable 300-1500 VDC|Regulation +-1% |
#2 High Voltage |
Adjustable 300-1500 VDC|Regulation +-1% |
Audio (click per event) |
On--Off switch on front panel |
Battery |
6 "AA" Alkaline-400 hours nominal |
Circuitry |
Digital Microprocessor Controlled |
Dimensions |
5.25" (14cm) H x 4" (10cm) W x 7" (18cm) L |
Weight |
2.5 lbs. (1.2 Kilos) w/Batteries |
Construction |
Rugged Heavy Duty Cast Aluminum & Stainless Steel |
Data Output (optional) |
9 Pin Serial RS-232 - Panel switch manual data extraction (optional) |