Highlights
Test procedure acc. IEC 61000-4-6 Ed. 4
Built-in amplifier saturation and linearity check
Built-in amplifier, 10 kHz to 400 MHz
Built-in power meter, 9 kHz to 1 GHz
Built-in dual directional coupler
Automatic calibration via built-in power meter
Connection for external amplifiers up to 1 GHz
USB-Interface
Software for Windows 7, Windows 8
Benefits:
One single box replaces five dedicated items. Hence setting up for testing is much easier and faster, saves space, is much more reliable and you achieve a higher reproducibility of the test results.
The CWS 500N1.3 offers the best benefit-to-cost ratio. Calibration/levelling of the test set-up has never been easier; terminate the DUT port with the calibration adapter, connect this to the 150ohm-to-50ohm matching impedance and lead its output to the Cal Input. Start the calibration routine and you are done in a minute.
The signal generator delivers signals up to 400MHz for the internal amplifier while the frequency range can be extended for 9kHz to 1GHz using an external amplifier.
The powerful, top-performance built-in class A amplifier allows to test up to 30V using any kind of coupling/decoupling device (CDNs, EM clamps, current injection clamps). The built-in 1GHz current monitor offers the ability to control the injected current signal while using injection clamps.