| ELECTRICAL |
| Voltage Output |
AC: 0 - 5000V
DC: 0 - 7000V |
| Max Leakage Current Output |
AC: 5000V = 200mA
DC: 5000V= 200mA, 7000V=140mA |
| Leakage Current Limit Adjustment |
AC: 5 – 200 mA
DC: 5 – 200 mA |
| Pass/Fail Criteria |
|
Leakage Current
|
Pass/Fail point user adjustable.
|
|
Dielectric Breakdown
|
Separate high frequency detection circuit for
breakdown spike detection
|
| Test Time |
User adjustable 1 to 90 sec. 1 sec steps, defeatable |
| Voltage Meter Accuracy AC |
±1.0% FSC from 0.50KV to 5.00KV volts
DC: ±1.0% FSC from 0.50KV to 7.00KV volts |
| Leakage Meter Accuracy |
± 1mA |
| Duty cycle |
100 % |
| Test adjustments |
Front Panel:
Test Time
Leakage Limit
Voltage Adjustment
Timer ON/DEFEAT
Hipot ON/DEFEAT
|
| ENVIRONMENTAL |
| Operating Temperature |
15-40 C |
| Relative Humidity Range |
0-90% non-condensing |
| GENERAL |
| Input power requirements |
208VAC, 3f, 50/60Hz, 6A/phase. |
| Weight |
90 lbs. |
| SAFETY AGENCY TOPICS |
| Transformer Output |
> 500VA |
| Visual Indication of Voltage Output |
Provided by front panel meter, directly connected to
high voltage output |
| Failure Indication |
Audible, provided by internal buzzer
Visual, provided by red LEDs, on front panel
Test can be automatically terminated on failure |
| Leakage Test |
Provided; 5 mA AC factory set pass/fail point, user
adjustable. |
| Zero Voltage Start |
It will NOT let to start the test until the voltage
adjustment knob is at the minimum position. |