Humidity |
Generation Range |
10-90% RH |
Accuracy
Control Element |
≤±1% RH (10–70% RH)
≤±1.5% RH (70–90% RH) |
Stability |
±0.2% RH (20–80% RH) |
Temperature |
Generation Range |
+10 to +50°C
(Lowest T set point = 10°C below ambient) |
Accuracy |
±0.1°C |
Stability |
±0.1°C |
Chamber |
Ramp Rate from
+20 to +40°C
+40 to +20°C |
1.5°C/minute
0.7°C minute |
Control Element |
Removable relative humidity sensor |
General |
Probe Ports |
up to 5 – sensor body diameters 5 to 25mm accommodated by port adapters |
Chamber Volume |
2000cm³ |
Chamber Dimensions |
105 x 105 x 160mm (h x w x d) |
Instrument Dimensions |
290 x 520 x 420mm (h x w x d) |
Set Point Resolution |
0.1 for humidity and temperature |
Displays |
3 digit LED, 10mm characters |
Supply |
85 to 264 V AC, 47/63 Hz, 150 VA |
Weight |
20kg |