Calibration measurement OCEANUS 6 AOIP

Mã sản phẩm: OCEANUS 6 |
Thương hiệu: AOIP

Giá bán: Liên hệ
Calibration measurement
Model: OCEANUS 6
Hãng: AOIP


Yêu Cầu Giá Tốt

Thông tin sản phẩm

The S model additionally includes an in-built temperature digital indicator to which an external temperature sensor can be connected, giving greater accuracy eliminating temperature gradient and loading errors.

For liquid bath, surface sensor and Blackbody use, an external thermometer should always be used. 

Oceanus 6 calibrates temperature probes from -25°C (in ambient of 20°C) to 140°C, with a substantially large calibration volume, 52 mm diameter by 300 mm deep. It can be used as a dry block, a liquid bath and as a blackbody source. As a traditional dry block, several thermometers can be quickly calibrated. Accessories can convert it to a stirred liquid bath, for surface sensor calibration, to calibrate infrared thermometers and even to be used as an ITS-90 fixed point system with calibration uncertainties as small as 0.0002°C.

Oceanus 6 is available in two models, the BASIC (B) and the SITE (S).

The B model includes a sophisticated temperature controller with a dual display for set temperature and dry block temperature,

For laboratory accuracy, Oceanus 6 can be used with high-end temperature indicator.

The Oceanus 6 offers unprecedented accuracies of 0.0002°C (2 sigma) at the Water Triple Point and the Gallium Melt temperature and up to 0.005°C in the stirred liquid bath option (by comparison).

Thông số kỹ thuật


 OCEANUS 6 thong minh


Sản phẩm cùng loại PEGASUS 1200 AOIP